Academic radiology · 2013
Removal of comfort pads underneath babies: a method of reducing radiation exposure to neonates.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Rattan Amit S, Cohen Mervyn D
Аннотация
RATIONALE AND OBJECTIVES: The recent increasing utilization of imaging has increased the population exposure to ionizing radiation. With increasing knowledge of the potential harm of radiation exposure, efforts should be made to minimize patient radiation whenever possible, especially in young children. The purpose of this study was to use the exposure index (EI) standard to assess the potential for reducing radiation dose to babies by removing a soft comfort pad, often placed underneath the baby. The pad is located between the baby and the image detector plate. As such it absorbs x-rays that have already passed through the baby but have not yet reached the imaging detector plate.
MATERIALS AND METHODS: Using a thoracic infant phantom and fixed exposure factors, we measured the percentage of the radiation exiting a neonatal chest phantom that was absorbed/attenuated by the comfort pad, before it hit the detector to create the image. We studied comfort pads of 4 different thicknesses, ranging from 0.5″ to 8″.
RESULTS: Radiation beam attenuation, ranging from 12% to 72.1%, was found with all comfort pads, with increased x-ray beam attenuation occurring with increasing pad thickness.
CONCLUSIONS: The study shows that comfort pads cause a high attenuation of the radiation beam, after it exits the chest phantom. As such, removal of the pads prior to radiographic exposure of babies is a method of potentially reducing patient radiation exposure in the newborn nursery.
MATERIALS AND METHODS: Using a thoracic infant phantom and fixed exposure factors, we measured the percentage of the radiation exiting a neonatal chest phantom that was absorbed/attenuated by the comfort pad, before it hit the detector to create the image. We studied comfort pads of 4 different thicknesses, ranging from 0.5″ to 8″.
RESULTS: Radiation beam attenuation, ranging from 12% to 72.1%, was found with all comfort pads, with increased x-ray beam attenuation occurring with increasing pad thickness.
CONCLUSIONS: The study shows that comfort pads cause a high attenuation of the radiation beam, after it exits the chest phantom. As such, removal of the pads prior to radiographic exposure of babies is a method of potentially reducing patient radiation exposure in the newborn nursery.