Does dual-energy CT of lower-extremity tendons incur penalties in patient radiation exposure or reduced multiplanar reconstruction image quality?

Lohan Derek G, Motamedi Kambiz, Chow Kira, Habibi Reza, Panknin Christoph, Ruehm Stefan G, Seeger Leanne L

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

The purposes of this study were to evaluate the quality and radiation exposure of data acquired with dual-energy CT compared with single-energy MDCT in the depiction of lower-extremity tendons and to assess whether a dual-energy CT voltage exists at which the quality of tendon depiction is optimal.

SUBJECTS AND

METHODS

Eleven healthy volunteers and seven clinically referred patients (10 men, eight women; mean age, 43.1 years; range, 20-71 years) underwent conventional single-energy CT and dual-energy CT examinations of both lower extremities with a dual-source CT scanner. Dual-energy reconstructions were made at combined tube voltages approximating 86, 98, 110, 122, and 134 kVp. Quantitative and qualitative analyses were performed on six tendons in each lower extremity, and the findings were compared with single-energy CT findings. The radiation dose involved was recorded in each case.

RESULTS

A trend toward increasing tendon attenuation was observed with increasing reconstructed tube voltage. The group of single-energy CT reconstructions proved significantly superior to each of the dual-energy CT reconstructions with regard to signal-to-noise ratio (F = 35.25, p < 0.0001) and contrast-to-noise ratio (F = 37.19, p < 0.0001), although interobserver agreement in subjective ranking was poor. Dual-energy CT had a significantly higher radiation dose (p < 0.05) than single-energy CT.

CONCLUSION

Dual-energy CT of lower-extremity tendons, irrespective of the reconstruction tube voltage chosen, yields multiplanar reformations inferior to those of single-energy CT with regard to signal-to-noise and contrast-to-noise ratios while involving significantly escalated patient exposure to ionizing radiation. Whether the tissue-differentiating promise of dual-energy CT is realized in future studies and warrants such concessions remains to be seen.

Для цитирования: Lohan Derek G, Motamedi Kambiz, Chow Kira, Habibi Reza, Panknin Christoph, Ruehm Stefan G, Seeger Leanne L. Does dual-energy CT of lower-extremity tendons incur penalties in patient radiation exposure or reduced multiplanar reconstruction image quality?. AJR. American journal of roentgenology. 2008. DOI: 10.2214/ajr.07.3624.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.