MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/s41440-023-01258-2

Медицинская библиотека · Hypertension research : official journal of the Japanese Society of Hypertension · 2023

Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: Preliminary results from a prospective single-center study.

Almeida Tiago P, Cortés Meritxell, Perruchoud David, Alexandre Jérémy, Vermare Pascale, Sola Josep, Shah Jay, Marques Luisa, Pellaton Cyril

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

In this preliminary study, we compared daytime blood pressure (BP) measurements performed by a commercially available cuffless-and continual-BP monitor (Aktiia monitor, Neuchâtel, Switzerland) and a traditional ambulatory BP monitor (ABPM; Dyasis 3, Novacor, Paris, France) from 52 patients enrolled in a 12-week cardiac rehabilitation (CR) program (Neuchâtel, Switzerland). Daytime (9am-9pm) systolic (SBP) and diastolic (DBP) BP from 7-day averaged data from Aktiia monitor were compared to 1-day averaged BP data from ABPM. No significant differences were found between the Aktiia monitor and the ABPM for SBP (μ ± σ [95% confidence interval]: 1.6 ± 10.5 [-1.5, 4.6] mmHg, P = 0.306; correlation [R2]: 0.70; ± 10/ ± 15 mmHg agreements: 60%, 84%). Marginally non-significant bias was found for DBP (-2.2 ± 8.0 [-4.5, 0.1] mmHg, P = 0.058; R2: 0.66; ±10/±15 mmHg agreements: 78%, 96%). These intermediate results show that daytime BP measurements using the Aktiia monitor generate data comparable to that of an ABPM monitor.

Для цитирования: Almeida Tiago P, Cortés Meritxell, Perruchoud David, Alexandre Jérémy, Vermare Pascale, Sola Josep, Shah Jay, Marques Luisa, Pellaton Cyril. Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: Preliminary results from a prospective single-center study.. Hypertension research : official journal of the Japanese Society of Hypertension. 2023. DOI: 10.1038/s41440-023-01258-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.