DOI 10.1002/mrm.29894

Медицинская библиотека · Magnetic resonance in medicine · 2023

Influence of patient head definition on induced E-fields during MR examination.

Goren Tolga, Reboux Sylvain, Farcito Silvia, Lloyd Bryn, Kuster Niels

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: Radiofrequency (RF) exposure during MR examination is limited by IEC 60601-2-33 to prevent thermal hazards to patients. These limits are also the basis to derive the maximum induced field for the demonstration of MR safety of implants per ISO/TS 10974 (2018). One limit is the head-averaged specific absorption rate (SAR), for which the head extent is defined differently by MR and implant vendors. The purpose of this technical note is to inform MR safety stakeholders on the sensitivity of safety evaluations due to different head extent definitions.

METHODS

RF distributions from the validated MRIxViP exposure libraries of 12 high-resolution human anatomical models were scaled to the normative SAR limits for different definitions of the head extent to compare the corresponding induced SAR and electric (E-)field levels.

RESULTS

The definitions of the head extent used by major implant vendors and defined in ISO/TS 10974 (2018) are larger than those introduced in IEC 60601-2-33 (2022), resulting in lower RF head exposure by up to 2.4 dB (factor 1.7). Other proposed definitions of the head result in intermediate values.

CONCLUSION

The different head extents result in different maximum RF exposures affecting the risk assessment by up to a factor of 1.7. The results of this study can be used to estimate the additional uncertainty in safety assessments. Future revisions of MR standards should eliminate this inconsistency.

Для цитирования: Goren Tolga, Reboux Sylvain, Farcito Silvia, Lloyd Bryn, Kuster Niels. Influence of patient head definition on induced E-fields during MR examination.. Magnetic resonance in medicine. 2023. DOI: 10.1002/mrm.29894.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.