MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1442-9071.2010.02471.x

Медицинская библиотека · Clinical & experimental ophthalmology · 2011

Efficacy and safety assessment of a novel ultraviolet C device for treating corneal bacterial infections.

Dean Simon J, Petty Alex, Swift Simon, McGhee Jennifer J, Sharma Anant, Shah Sunil, Craig Jennifer P

СтатьяВалидационное исследованиеПеревод в очереди

Аннотация

BACKGROUND

A prototype solid-state Ultraviolet-C (UVC) LED device may be useful in the treatment of corneal microbial infections, as UVC is commonly used for eradicating bacteria, fungi and viruses in other settings. This study assessed the efficacy of 265 nm UVC from this LED, on four different bacterial strains, and investigated the consequences of corresponding exposures on human corneal epithelial cells in vitro.

METHODS

Agar plate lawns of Staphylococcus aureus, Escherichia coli, Pseudomonas aeruginosa and Streptococcus pyogenes were exposed to a 4.5 mm diameter 265 nm UVC beam at a fixed intensity and distance, for 30, 5, 4, 2 and 1 seconds. Growth inhibition was assessed with a BioRad Gel imager, and the diameter of lucent areas of bacterial inhibition recorded. Human corneal epithelial cells cultured on glass cover-slips were exposed to corresponding doses of UVC from the same device. Live/dead staining was performed and the results quantified.

RESULTS

There was 100% inhibition of growth for all bacteria tested, at all exposure times. A 30-second exposure of human corneal epithelium to UVC gave no statistically significant decrease (P = 0.877) in the ratio of live to dead cells when compared to control cultures.

CONCLUSION

The results confirmed that a 1 second exposure to germicidal UVC from this LED source was sufficient to inhibit microbial proliferation in the four bacterial strains tested in vitro. The literature suggests UVC at this dose could potentially be beneficial in treating corneal surface infections, without causing significant adverse effects, supported by our findings in human corneal epithelium exposed to UVC.

Для цитирования: Dean Simon J, Petty Alex, Swift Simon, McGhee Jennifer J, Sharma Anant, Shah Sunil, Craig Jennifer P. Efficacy and safety assessment of a novel ultraviolet C device for treating corneal bacterial infections.. Clinical & experimental ophthalmology. 2011. DOI: 10.1111/j.1442-9071.2010.02471.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.