Accuracy of wearable heart rate monitors in cardiac rehabilitation.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Etiwy Muhammad, Akhrass Zade, Gillinov Lauren, Alashi Alaa, Wang Robert, Blackburn Gordon, Gillinov Stephen M, Phelan Dermot, Gillinov A Marc, Houghtaling Penny L, Javadikasgari Hoda, Desai Milind Y
Аннотация
BACKGROUND: To assess the accuracy of four wearable heart rate (HR) monitors in patients with established cardiovascular disease enrolled in phase II or III cardiac rehabilitation (CR).
METHODS: Eighty adult patients enrolled in phase II or III CR were monitored during a CR session that included exercise on a treadmill and/or stationary cycle. Participants underwent HR monitoring with standard ECG limb leads, an electrocardiographic (ECG) chest strap monitor (Polar H7), and two randomly assigned wrist-worn HR monitors (Apple Watch, Fitbit Blaze, Garmin Forerunner 235, TomTom Spark Cardio), one on each wrist. HR was recorded at rest and at 3, 5, and 7 minutes of steady-state exercise on the treadmill and stationary cycle.
RESULTS: Across all exercise conditions, the chest strap monitor (Polar H7) had the best agreement with ECG (rc=0.99) followed by the Apple Watch (rc=0.80), Fitbit Blaze (rc=0.78), TomTom Spark (rc=0.76) and Garmin Forerunner (rc=0.52). There was variability in accuracy under different exercise conditions. On the treadmill, only the Fitbit Blaze performed well (rc=0.76), while on the stationary cycle, Apple Watch (rc=0.89) and TomTom Spark (rc=0.85) were most accurate.
CONCLUSIONS: In cardiac patients, the accuracy of wearable, optically based HR monitors varies, and none of those tested was as accurate as an electrode-containing chest monitor. This observation has implications for in-home CR, as electrode-containing chest monitors should be used when accurate HR measurement is imperative.
METHODS: Eighty adult patients enrolled in phase II or III CR were monitored during a CR session that included exercise on a treadmill and/or stationary cycle. Participants underwent HR monitoring with standard ECG limb leads, an electrocardiographic (ECG) chest strap monitor (Polar H7), and two randomly assigned wrist-worn HR monitors (Apple Watch, Fitbit Blaze, Garmin Forerunner 235, TomTom Spark Cardio), one on each wrist. HR was recorded at rest and at 3, 5, and 7 minutes of steady-state exercise on the treadmill and stationary cycle.
RESULTS: Across all exercise conditions, the chest strap monitor (Polar H7) had the best agreement with ECG (rc=0.99) followed by the Apple Watch (rc=0.80), Fitbit Blaze (rc=0.78), TomTom Spark (rc=0.76) and Garmin Forerunner (rc=0.52). There was variability in accuracy under different exercise conditions. On the treadmill, only the Fitbit Blaze performed well (rc=0.76), while on the stationary cycle, Apple Watch (rc=0.89) and TomTom Spark (rc=0.85) were most accurate.
CONCLUSIONS: In cardiac patients, the accuracy of wearable, optically based HR monitors varies, and none of those tested was as accurate as an electrode-containing chest monitor. This observation has implications for in-home CR, as electrode-containing chest monitors should be used when accurate HR measurement is imperative.