Exposure of workers to pulsed gradients in MRI.

Crozier Stuart, Wang Hua, Trakic Adnan, Liu Feng

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To numerically evaluate the electric field/current density magnitudes and spatial distributions in healthcare workers when they are standing close to the gradient coil windings near the magnetic resonance imaging (MRI) scanner ends.

MATERIALS AND

METHODS

Anatomically realistic, whole-body male and female voxel phantoms are engaged to model the workers at various positions near the ends of three cylindrical gradient coils (x-, y-, and z-axis gradients). The numerical calculations of induced fields are based on an efficient, quasistatic finite-difference method.

RESULTS

The simulations show that it is possible to induce electric fields/current densities above levels recommended by the International Commission for Non-ionizing Radiation Protection (ICNIRP) and the Institute of Electrical and Electronics Engineers (IEEE) standards when the workers are standing close to the gradient coils and when two or three gradients are switched simultaneously, as is often the case.

CONCLUSION

The longitudinal gradient tends to induce more fields in workers than the transverse coils. The strongest levels of field exposure are observed when all three gradients are operated simultaneously and can be above regulations when the healthcare worker is close to the gradient coils. Other postures such as bending into the magnet shall be investigated in further studies.

Для цитирования: Crozier Stuart, Wang Hua, Trakic Adnan, Liu Feng. Exposure of workers to pulsed gradients in MRI.. Journal of magnetic resonance imaging : JMRI. 2007. DOI: 10.1002/jmri.21162.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.