m+Medelevator.RU
← Поиск публикаций
Annals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc · 2015

R/S Ratio Variability in Lead V1 Observed between Selected Four 1-Minute ECG Fragments of 24-Hour ECG as a Predictor of Incomplete Resynchronization during Full 24-Hour ECG: Pilot Study.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Pyszno-Prokopowicz Dominika, Baranowski Rafał, Bodalski Robert, Madej Magdalena, Piotrowicz Ryszard

Аннотация

BACKGROUND: About 30% of patients do not have an effective cardiac resynchronization therapy (CRT). Routine assessment of CRT by devices interrogation (DI) is not entirely reliable. Additional information provide detailed QRS analysis in 24-hour ECG, however it is time-consuming. The aim of the study was the assessment of R/S ratio variability in lead V1 between selected fragments of 24-hour ECG as a predictor of incomplete biventricular pacing (BIVP) during full 24-hour ECG.

METHODS: The 12-lead 24-hour ECGs of 43 patients with sinus rhythm were studied. During 24-hour ECG the 6-minute walk test (6MWT) was performed. The CRT was assessed by analyzing DI and full 24-hour ECG and four 1-minute fragments of these ECG recordings: during the minimum and the maximum heart rate and at the 1st and last minute of 6MWT.

RESULTS: During DI the effective (>95%) BIVP was present in 36 patients (83.7%). Analysis of full 24-hour ECG confirmed appropriate BIVP in 31 patients (72%) and suspected incomplete BIVP (≤95%) in 12 patients (28%). In 9/12 patients the R/S ratio variability in lead V1 was visible between selected ECG fragments of 24-hour ECG. These results were not associated with the results of DI but were significantly associated with full 24-hour analysis of QRS.

CONCLUSIONS: R/S variability in lead V1 between selected fragments of 24-hour ECG can be considered a predictor of potentially incomplete BIVP confirmed by further complete 24-hour ECG analysis in patients with appropriate pacing reported during DI.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты