Endoscopy international open · 2025
Enhancing emergency endoscopy efficiency with an additional suction channel: In vitro assessment.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sasse André, Heiduk Thomas Roland, Scheunemann Marvin, Hiebel Lukas, Knoop Richard F, Adler Marius, Seif Amir Hosseini Ali, Wedi Edris, Amanzada Imke, Ellenrieder Volker, Petzold Golo, Amanzada Ahmad
Аннотация
BACKGROUND AND STUDY AIMS: Emergency endoscopic interventions for upper gastrointestinal hemorrhage are frequently hampered by presence of blood clots and food debris. This study aimed to assess whether integration of an additional suction channel (ASC) onto a standard gastroscope enhances efficiency of aspirating clots and viscous fluids.
PATIENTS AND METHODS: A 5.3-mm suction catheter was used as an ASC mounted on a 2.8-mm standard gastroscope. Suction efficacy was evaluated using gastroscopes with working channel diameters of 2.8 mm, 3.7 mm, and 6 mm in vitro. Defined volumes of water, fruit yogurt, and coagulated blood were aspirated, and time required for complete evacuation was measured. Each setup was tested with and without the BioVac system.
RESULTS: The ASC significantly enhanced suction performance across all test media. Notably, the 2.8-mm gastroscope with ASC outperformed all other configurations in aspirating water and yogurt. For clotted blood, the ASC significantly improved evacuation times compared with all other setups besides 6-mm + BioVac.
CONCLUSIONS: A standard gastroscope equipped with an ASC significantly enhances suction performance in an in vitro model, outperforming gastroscopes with larger working channels. These findings warrant further validation in an ex vivo model to determine their clinical applicability.
PATIENTS AND METHODS: A 5.3-mm suction catheter was used as an ASC mounted on a 2.8-mm standard gastroscope. Suction efficacy was evaluated using gastroscopes with working channel diameters of 2.8 mm, 3.7 mm, and 6 mm in vitro. Defined volumes of water, fruit yogurt, and coagulated blood were aspirated, and time required for complete evacuation was measured. Each setup was tested with and without the BioVac system.
RESULTS: The ASC significantly enhanced suction performance across all test media. Notably, the 2.8-mm gastroscope with ASC outperformed all other configurations in aspirating water and yogurt. For clotted blood, the ASC significantly improved evacuation times compared with all other setups besides 6-mm + BioVac.
CONCLUSIONS: A standard gastroscope equipped with an ASC significantly enhances suction performance in an in vitro model, outperforming gastroscopes with larger working channels. These findings warrant further validation in an ex vivo model to determine their clinical applicability.