MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2106/jbjs.26.00470

Медицинская библиотека · The Journal of bone and joint surgery. American volume · 2026

Antibiotic-Loaded Ceramic Bone Grafts Exhibit Burst, Not Sustained, Gentamicin Release: In Vivo Evidence and Reanalysis of in Vitro Assay Artifacts Informing Clinical Use.

Berthelsen Catrine Jyde, Crocoll Christoph, Poborsky Michal, Jensen Louise Kruse, Henriksen Jonas Rosager, Hansen Anders Elias, Bue Mats

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Antibiotic-loaded ceramic bone grafts (ACBGs) are widely assumed to provide extended local antimicrobial protection, maintaining concentrations above the minimum inhibitory concentration (MIC) for 28 to 42 days. However, this time frame is mainly based on in vitro elution assays with simplified washout models, and its relevance to in vivo performance is uncertain. This study compared the in vivo and in vitro release and behavior of gentamicin from 2 commercial ACBGs.

METHODS

Gentamicin-loaded hydroxyapatite/calcium sulfate (HA-CS) and calcium sulfate (CS) composites were investigated using a murine implantation model and in vitro elution assays. Gentamicin content in recovered depot material and elution media was quantified by liquid chromatography with tandem mass spectrometry. Model calculations evaluated how the assay design affects the estimated time above the MIC.

RESULTS

Both ACBGs released >98% of gentamicin within 6 hours and >99% within 24 hours in vivo. In vitro assays likewise indicated early depot depletion, with concentration-time profiles beyond 48 hours driven by serial dilution rather than release. Model calculations suggested that the apparent prolonged antibiotic activity in previous in vitro assays was largely due to the assay design, not to sustained release, implying that freely administered antibiotics could produce similar profiles.

CONCLUSIONS

Neither ACBG demonstrated measurable sustained release beyond an initial 24-hour burst phase. This may have important clinical implications, as expectations of prolonged local antimicrobial protection are not supported by the cumulative-release data showing rapid antibiotic depot depletion.

Для цитирования: Berthelsen Catrine Jyde, Crocoll Christoph, Poborsky Michal, Jensen Louise Kruse, Henriksen Jonas Rosager, Hansen Anders Elias, Bue Mats. Antibiotic-Loaded Ceramic Bone Grafts Exhibit Burst, Not Sustained, Gentamicin Release: In Vivo Evidence and Reanalysis of in Vitro Assay Artifacts Informing Clinical Use.. The Journal of bone and joint surgery. American volume. 2026. DOI: 10.2106/jbjs.26.00470.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.