DOI 10.1177/0363546519826953

Медицинская библиотека · The American journal of sports medicine · 2019

Evaluation of an In-Ear Sensor for Quantifying Head Impacts in Youth Soccer.

Sandmo Stian Bahr, McIntosh Andrew S, Andersen Thor Einar, Koerte Inga Katharina, Bahr Roald

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Wearable sensor systems have the potential to quantify head kinematic responses of head impacts in soccer. However, on-field use of sensors (eg, accelerometers) remains challenging, owing to poor coupling to the head and difficulties discriminating low-severity direct head impacts from inertial loading of the head from human movements, such as jumping and landing.

PURPOSE: To test the validity of an in-ear sensor for quantifying head impacts in youth soccer.

STUDY

DESIGN

Descriptive laboratory study.

METHODS

First, the sensor was mounted to a Hybrid III headform and impacted with a linear impactor or a soccer ball. Peak linear acceleration (PLA), peak rotational acceleration (PRA), and peak rotational velocity (PRV) were obtained from both systems; random and systematic errors were calculated with Hybrid III as reference. Then, 6 youth soccer players wore sensors and performed a structured training protocol, including heading and nonheading exercises; they also completed 2 regular soccer sessions. For each accelerative event recorded, PLA, PRA, and PRV outputs were compared with video recordings. Receiver operating characteristic curves were used to determine the sensor's discriminatory capacity in both on-field settings, establishing cutoff values for predicting outcomes.

RESULTS

For the laboratory tests, the random error was 11% for PLA, 20% for PRA, and 5% for PRV; the systematic error was 11%, 19%, and 5%, respectively. For the structured training protocol, heading events resulted in higher absolute values (PLA = 15.6 g± 11.8 g) than nonheading events (PLA = 4.6 g± 1.2 g); the area under the curve was 0.98 for PLA. In regular training sessions, the area under the curve was >0.99 for PLA. A 9 g cutoff value yielded a positive predictive value of 100% in the structured training protocol versus 65% in the regular soccer sessions.

CONCLUSION

The in-ear sensor displayed considerable random error and substantially overestimated head impact exposure. Despite the sensor's excellent on-field accuracy for discriminating headings from other accelerative events in youth soccer, absolute values must be interpreted with caution, and there is a need for secondary means of verification (eg, video analysis) in real-life settings.

CLINICAL RELEVANCE: Wearable sensor systems can potentially provide valuable insights into head impact exposures in contact sports, but their limitations require careful consideration.

Для цитирования: Sandmo Stian Bahr, McIntosh Andrew S, Andersen Thor Einar, Koerte Inga Katharina, Bahr Roald. Evaluation of an In-Ear Sensor for Quantifying Head Impacts in Youth Soccer.. The American journal of sports medicine. 2019. DOI: 10.1177/0363546519826953.