DOI 10.1111/dom.70404

Медицинская библиотека · Diabetes, obesity & metabolism · 2025

In-flight glycaemic control using automated insulin delivery systems: A within-subject comparative pilot study.

Meçani Renald, Basta Silvia, Baumann Petra M, Cigler Monika, El Hakouni Omaima, Hochfellner Daniel A, Shojaee-Moradie Fariba, Fan Ka Siu, Tews Dietrich, Pieber Thomas R, Russell-Jones David, Mathieu Chantal, Koehler Gerd, Mader Julia K

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Automated insulin delivery systems (AID) are increasingly used by people living with type 1 diabetes, but their functionality during air travel remains insufficiently studied. Some concerns have arisen regarding safety using AID systems under flight conditions, particularly related to potential insulin delivery errors induced by cabin pressure changes.

METHODS

This study assessed the safety and efficacy commercial AIDs in 20 adults with type 1 diabetes (age: 37.9 ± 12.8 years, HbA1c: 52.3 ± 4.9 mmol/mol, TIR: 74 ± 9.8%, diabetes duration: 19.25 ± 10.6 years) during mid-haul commercial flights and matched ground-based conditions (MiniMed780G; Omnipod 5; Control-IQ; and CamAPS FX with YpsoPump, n = 5 per system). Outcomes included time in range (TIR, 3.9-10.0 mmol/L), time in tight range (TITR, 3.9-7.8 mmol/L), time above range 1 (TAR1, 10.1-13.9 mmol/L), time above range 2 (TAR2, >13.9 mmol/L), time below range 1 (TBR1, 3.0-3.8 mmol/L), time below range 2 (TBR2, <3.0 mmol/L), and total daily insulin doses.

RESULTS

Overall TIR was 88.2 ± 13.5% in-flight versus 82.1 ± 17.7% on-ground. The mean difference in TIR was 4.6%, (95% CI: -5.0- 14.2). Other overall metrics (in-flight versus on-ground) were: TITR = 67.7 ± 19.4% vs. 64.2 ± 18.1%, TAR = 7.1 ± 10.5% vs. 12.0 ± 13.8% and TBR = 4.7 ± 7.2% vs. 5.9 ± 15.7%. Insulin delivery was slightly increased in-flight (25.4 ± 17.6 IU vs. 18.7 ± 11.6 IU), driven by bolus dosing (17.8 ± 13.3 IU vs. 11.9 ± 6.6 IU). No episodes of level 2 hypoglycaemia, diabetic ketoacidosis, or device failure occurred.

CONCLUSIONS

AID systems maintained robust glycaemic control and safety during commercial flights, comparable to ground-based conditions. These data support the feasibility of in-flight AIDs use and provide a basis for future regulatory and clinical practice updates. Larger, multicenter studies are needed to generalize findings.

Для цитирования: Meçani Renald, Basta Silvia, Baumann Petra M, Cigler Monika, El Hakouni Omaima, Hochfellner Daniel A, Shojaee-Moradie Fariba, Fan Ka Siu, Tews Dietrich, Pieber Thomas R, Russell-Jones David, Mathieu Chantal, Koehler Gerd, Mader Julia K. In-flight glycaemic control using automated insulin delivery systems: A within-subject comparative pilot study.. Diabetes, obesity & metabolism. 2025. DOI: 10.1111/dom.70404.

Предметные рубрики MeSH