Operator safety during the acquisition of intraoral images with a handheld and portable X-ray device.
Аннотация
OBJECTIVES
The present study aims at investigating different radiation protection issues and dose values while acquiring intraoral images with a handheld X-ray device.
METHODS
An Aribex NOMAD Pro 2™, a RANDO® male head phantom, a consistency testing body, a PTW NOMEX® Multimeter, and a PTW Farmer® Ionization Chamber Type 30,010 were used to investigate: (1) dose area products; (2) the expansion of the control area (CA); (3) the scattering pattern and (4) the potential risk for operators of the X-ray device.
RESULTS
Dose area products at different exposure times were distributed linearly with a high correlation factor (>0.9). At 4000 simulated exposures, the greatest extent of the CA was 42 cm (mean = 16.7 cm, SD = 10.8 cm). The highest occurrence of scattering radiation resulted between the RANDO® phantom and the X-ray device. No scattered radiation was measured at the dorsal part of the phantom or on the operator site of a virtual vertical plane through the focal spot of the X-ray.
CONCLUSIONS
Through this study, we could demonstrate that the application of an Aribex NOMAD Pro 2 device for intraoral imaging does not increase the risk for the operator if the device is controlled according to the manufacturer's specifications. Furthermore, we were able to show that the CA was significantly smaller than specified by European and other international radiation protection standards.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.