MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/ajcp/aqaa117

Медицинская библиотека · American journal of clinical pathology · 2021

A Flow Cytometric Study of Reagent Cells to Resolve ABO Typing Discrepancy.

Maracaja Danielle L V, Qiao Jesse, Salazar Tomas, Barry James, LaForce Karen, Holder Kenneth, Olson John D

Клинический случайСтатьяПеревод в очереди

Аннотация

OBJECTIVES

RBC alloantibodies can lead to ABO grouping discrepancies unrelated to A or B antigens or antibodies posing challenges in the blood bank testing. Routine blood bank testing and flow cytometry were used to immunophenotype reagent cells and elucidate the cause of ABO discrepancies in two patients.

METHODS

ABO discrepancy was identified in two patients after transfusion with several units of RBCs. For both patients, the pretransfusion type and screen demonstrated blood group A. Eight and 16 days later, both patients showed an apparent antibody to reagent group A cells, which prompted additional study with patients' samples and flow cytometric testing of commercial reagent cells.

RESULTS

In both patients' specimens, posttransfusion evaluation demonstrated an emerging antibody to the Kell antigen (K). The RBCs of both patients typed negative for K, and both were transfused with K-positive RBCs. Flow cytometric analysis of reagent RBCs demonstrated that five of seven lot numbers were positive for K.

CONCLUSIONS

Emerging anti-K antibody led to agglutination of the K-positive reagent A1 cells, highlighting the importance of considering RBC alloantibodies and the composition of reagent cells when interpreting cases with an apparent ABO grouping discrepancy.

Для цитирования: Maracaja Danielle L V, Qiao Jesse, Salazar Tomas, Barry James, LaForce Karen, Holder Kenneth, Olson John D. A Flow Cytometric Study of Reagent Cells to Resolve ABO Typing Discrepancy.. American journal of clinical pathology. 2021. DOI: 10.1093/ajcp/aqaa117.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.