DOI 10.1016/j.jcv.2014.03.009

Медицинская библиотека · Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology · 2014

Comparison of two automated immunoassays for the determination of Puumalavirus IgM and IgG.

Muyldermans Astrid, Lagrou Katrien, Patteet Sofie, Van Esbroeck Marjan, Van Ranst Marc, Saegeman Veroniek

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Puumala virus (PUUV), a member of the genus hantavirus, can cause nephropathia epidemica, a mild form of haemorrhagic fever with renal syndrome. The method of choice for the serodiagnosis of hantavirus infections are enzyme-linked immunosorbent assays (ELISAs).

OBJECTIVES

Two commercially available PUUV ELISA kits were compared: Hantavirus (Puumala) IgM/IgG ELISA (Progen, Heidelberg, Germany) and PUUMALA IgM and IgG EIA AutoM (Reagena, Toivala, Finland).

STUDY

DESIGN

The sensitivity of the ELISA kits was evaluated with a panel of 55 serum samples from patients with an acute (n=27) or past (n=28) infection based on Progen or Reagena. A panel of 56 serum samples was composed to evaluate the specificity: samples with potentially false positive Progen Puumala IgM results (n=12), seronegative samples for Puumala IgG/IgM with Progen (n=20), and potentially cross reacting samples (n=24). Discrepancies between the two assays were resolved with strip immunoblot. As measure of agreement between Progen and Reagena results, Cohen kappa coefficient was calculated.

RESULTS

Reagena showed a higher specificity (IgM 100%, IgG 100%) than Progen Puumala (IgM 73.21%, IgG 100%). However, Reagena showed a slightly lower sensitivity (IgM 96.15%, IgG 97.78%) compared with Progen (IgM 100%, IgG 100%). Substantial agreement with a Cohen kappa of 0.67 and 0.76 was found between the two assays for Puumala IgM and IgG respectively.

CONCLUSIONS

This study showed a higher specificity of Reagena in comparison to Progen with a lower sensitivity, probably caused by selection bias. In spite of Reagena's lower sensitivity, no acute infection was missed with this assay.

Для цитирования: Muyldermans Astrid, Lagrou Katrien, Patteet Sofie, Van Esbroeck Marjan, Van Ranst Marc, Saegeman Veroniek. Comparison of two automated immunoassays for the determination of Puumalavirus IgM and IgG.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2014. DOI: 10.1016/j.jcv.2014.03.009.