BACKGROUND

The transmission of pathogens, antibodies, and proteins is a possible consequence of blood product transfusion. A female patient had an unexpected positive serum β-human chorionic gonadotropin result, indicative of pregnancy, after she had received a transfusion with 1 unit of platelet concentrate, 4 units of red blood cells, and 4 units of pooled solvent/detergent-treated plasma (Octaplas).

STUDY DESIGN AND

METHODS

To investigate the possibility of passive transfusion of β-human chorionic gonadotropin from the plasma transfusion, one additional unit from the same batch was thawed and analyzed. To validate the β-human chorionic gonadotropin assay for use in solvent/detergent-treated plasma and to investigate any interference in the assay, dilution experiments were performed using the implicated plasma batch diluted with male and non-pregnant female sera. Also, plasma from a known pregnant woman was diluted with Octaplas (tested negative for β-human chorionic gonadotropin) and with a male serum to validate the assay for use in solvent/detergent-treated plasma.

RESULTS

The implicated solvent/detergent-treated plasma had a mean β-human chorionic gonadotropin level of 91.5 mIU/mL. Results from the dilution experiments revealed an excellent correlation (r > 0.99) between β-human chorionic gonadotropin measurement in solvent/detergent-treated plasma and male serum and no over or under recovery of the expected results. Further measurements of β-human chorionic gonadotropin levels in the female recipient revealed an estimated half-life of 6 hours.

CONCLUSION

This case demonstrates the importance of considering the possibility of passive transmission of analytes to a patient from the transfusion of blood products. Furthermore, the measurement of β-human chorionic gonadotropin is valid in solvent/detergent-treated plasma using a Roche Cobas analyzer.

Для цитирования: Jilma-Stohlawetz Petra, Wreford-Bush Tim, Mills Francesca, Davidson Fiona, Kursten Friedrich W, Jilma Bernd, Birchall Janet. False-positive pregnancy test after transfusion of solvent/detergent-treated plasma.. Transfusion. 2017. DOI: 10.1111/trf.14304.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.