DOI 10.1016/j.cca.2017.02.013

Медицинская библиотека · Clinica chimica acta; international journal of clinical chemistry · 2017

Impact of technical and assay variation on reporting of hemolysis in stored red blood cell products.

Almizraq Ruqayyah J, Yi Qi Long, Acker Jason P, Biomedical Excellence for Safer Transfusion (BEST) Collaborative

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Hemolysis of RBCs is an important measure of product quality and is influenced by donor factors, blood component manufacturing and storage. Percent hemolysis is determined using hematocrit (Hct), supernatant Hb (SHb) and total Hb (THb), each of which can be measured using a variety of methods.

METHODS

Sixteen members of an international collaborative were surveyed to understand equipment and procedural variation in hemolysis testing. In a laboratory-based evaluation, we examined how hemolysis was impacted by: measurement of Hct, SHb, THb and number and force of centrifugations for SHb preparation. The number and size of extracellular vesicles (EVs) was also examined.

RESULTS

There was no consensus in equipment or procedures used by international laboratories to measure hemolysis. The centrifugation force used to prepare samples influenced SHb concentration when a single or double (p=0.0001) centrifugation step was used. The number and force of centrifugation related directly to the ability to remove EVs and EV-bound Hb from samples. Hemolysis varied significantly from 0.16% to 0.32% (mean of 0.22%) depending on the combination of methods or centrifugation conditions used to test expired samples.

CONCLUSION

Method and preparative procedures have a critical impact on measurement of hemolysis in RCC.

Для цитирования: Almizraq Ruqayyah J, Yi Qi Long, Acker Jason P, Biomedical Excellence for Safer Transfusion (BEST) Collaborative. Impact of technical and assay variation on reporting of hemolysis in stored red blood cell products.. Clinica chimica acta; international journal of clinical chemistry. 2017. DOI: 10.1016/j.cca.2017.02.013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.