BACKGROUND

The assessment of the coagulation status using thromboelastography (TEG) in Chinese population has less been reported. This study aimed to establish reliable reference values for kaolin-activated TEG in Chinese volunteers.

METHODS

A total of 1681 Chinese adult individuals were recruited for this study. The reference individuals were stratified by gender and age, and the TEG values were measured on the basis of strict quality control. The 95% reference values were determined using nonparametric statistical methods.

RESULTS

The sex-related 95% reference values were reaction time (R):4.2-8.7 minutes; clotting time (K): 1.2-3.2 minutes; alpha angle (α): 47.0-72.3 degree; maximum amplitude (MA): 49.1-70.5 mm for males, and R: 3.7-9.0 minutes; K: 1.0-3.2 minutes; α: 48.4-74.4 degree; MA: 46.8-72.4 mm for females. Also, the TEG parameters indicated a relatively more hypercoagulable profile in both female and elder groups.

CONCLUSIONS

This study established the reference values for kaolin-activated TEG in the target Chinese population, which might provide a reference for both clinical and laboratory studies.

Для цитирования: Sun Jun-Bo, Bian Mao-Hong, Zhong Tao, Lu Ying-Yu, Zhu Bang-Qiang, Wen Hui-Qin, Hu Hai-Liang. Reference values for kaolin-activated thromboelastography in volunteers of Anhui Province in China.. Journal of clinical laboratory analysis. 2017. DOI: 10.1002/jcla.22128.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.