Annals of laboratory medicine · 2017

Comparative Analysis of Clinical Samples Showing Weak Serum Reaction on AutoVue System Causing ABO Blood Typing Discrepancies.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Jo Su Yeon, Lee Ju Mi, Kim Hye Lim, Sin Kyeong Hwa, Lee Hyeon Ji, Chang Chulhun Ludgerus, Kim Hyung Hoi

Аннотация

BACKGROUND: ABO blood typing in pre-transfusion testing is a major component of the high workload in blood banks that therefore requires automation. We often experienced discrepant results from an automated system, especially weak serum reactions. We evaluated the discrepant results by the reference manual method to confirm ABO blood typing.

METHODS: In total, 13,113 blood samples were tested with the AutoVue system; all samples were run in parallel with the reference manual method according to the laboratory protocol.

RESULTS: The AutoVue system confirmed ABO blood typing of 12,816 samples (97.7%), and these results were concordant with those of the manual method. The remaining 297 samples (2.3%) showed discrepant results in the AutoVue system and were confirmed by the manual method. The discrepant results involved weak serum reactions (<2+ reaction grade), extra serum reactions, samples from patients who had received stem cell transplants, ABO subgroups, and specific system error messages. Among the 98 samples showing ≤1+ reaction grade in the AutoVue system, 70 samples (71.4%) showed a normal serum reaction (≥2+ reaction grade) with the manual method, and 28 samples (28.6%) showed weak serum reaction in both methods.

CONCLUSIONS: ABO blood tying of 97.7% samples could be confirmed by the AutoVue system and a small proportion (2.3%) needed to be re-evaluated by the manual method. Samples with a 2+ reaction grade in serum typing do not need to be evaluated manually, while those with ≤1+ reaction grade do.