Fully automated, clinical-grade bone marrow processing: a single-centre experience.
Аннотация
BACKGROUND
Clinical grade processing of harvested bone marrow is required in various clinical situations, particularly in the management of ABO mismatching in allogeneic haematopoietic stem cell transplantation (HSCT) and in regenerative medicine.
MATERIAL AND
METHODS
We report a single-centre experience using a fully automated, clinical grade, closed system (Sepax, Biosafe, Switzerland). From 2003 to 2015, 125 procedures were performed in our laboratory, including buffy-coat production for HSCT (n=58), regenerative medicine in an orthopaedic setting (n=54) and density-gradient separation in a trial for treatment of critical limb ischaemia (n=13).
RESULTS
Buffy coat separation resulted in a median volume reduction of 85% (range, 75-87%), providing satisfactory red blood cell depletion (69%, range 30-88%) and a median recovery of CD34 cells of 96% (range, 81-134%) in the setting of allogeneic HSCT. Significantly greater volume reduction (90%; range, 90-92%) and red blood cell depletion (88%; range, 80-93%) were achieved by the new SmartRedux software released for Sepax2, validated in the last eight allogeneic HSCT. The density gradient separation programme resulted in complete red blood cell depletion associated with high CD34 recovery (69%; range, 36-124%). No reactions related to the quality of the product were reported. Time to engraftment following allogeneic HSCT was in the normal range. No cases of microbiological contamination related to the manipulation were reported.
DISCUSSION: Clinical grade, automated bone marrow manipulation with Sepax was shown to be effective, giving operator-independent results and could be used for a broad range of clinical applications.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.