Air Quality Monitoring During High-Level Biocontainment Ground Transport: Observations From Two Operational Exercises.
Аннотация
OBJECTIVE
Stretcher transport isolators provide mobile, high-level biocontainment outside the hospital for patients with highly infectious diseases, such as Ebola virus disease. Air quality within this confined space may pose human health risks.
METHODS
Ambient air temperature, relative humidity, and CO2 concentration were monitored within an isolator during 2 operational exercises with healthy volunteers, including a ground transport exercise of approximately 257 miles. In addition, failure of the blower unit providing ambient air to the isolator was simulated. A simple compartmental model was developed to predict CO2 and H2O concentrations within the isolator.
RESULTS
In both exercises, CO2 and H2O concentrations were elevated inside the isolator, reaching steady-state values of 4434 ± 1013 ppm CO2 and 22 ± 2 mbar H2O in the first exercise and 3038 ± 269 ppm CO2 and 20 ± 1 mbar H2O in the second exercise. When blower failure was simulated, CO2 concentration exceeded 10 000 ppm within 8 minutes. A simple compartmental model predicted CO2 and H2O concentrations by accounting for human emissions and blower air exchange.
CONCLUSIONS
Attention to air quality within stretcher transport isolators (including adequate ventilation to prevent accumulation of CO2 and other bioeffluents) is needed to optimize patient safety.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.