IMPORTANCE

Devices enabled by artificial intelligence (AI) and machine learning (ML) are increasingly used in clinical settings, but there are concerns regarding benefit-risk assessment and surveillance by the US Food and Drug Administration (FDA).

OBJECTIVE

To characterize pre- and postmarket efficacy, safety, and risk assessment reporting for FDA-cleared AI/ML devices.

DESIGN AND

SETTING

This was a cross-sectional study using linked data from FDA decision summaries and approvals databases, the FDA Manufacturer and User Facility Device Experience Database, and the FDA Medical Device Recalls Database for all AI/ML devices cleared by the FDA from September 1995 to July 2023. Data were analyzed from October to November 2024.

MAIN OUTCOMES AND MEASURES: AI/ML reporting of study design, data availability, efficacy, safety, bias assessments, adverse events, device recalls, and risk classification.

RESULTS

The analysis included data for all 691 AI/ML devices that received FDA clearance through 2023, with 254 (36.8%) cleared in or after 2021. Device summaries often failed to report study designs (323 [46.7%]), training sample size (368 [53.3%]), and/or demographic information (660 [95.5%]). Only 6 devices (1.6%) reported data from randomized clinical trials and 53 (7.7%) from prospective studies. Few premarket summaries contained data published in peer-reviewed journals (272 [39.4%]) or provided statistical or clinical performance, including sensitivity (166 [24.0%]), specificity (152 [22.0%]), and/or patient outcomes (3 [<1%]). Some devices reported safety assessments (195 [28.2%]), adherence to international safety standards (344 [49.8%]), and/or risks to health (42 [6.1%]). In all, 489 adverse events were reported involving 36 (5.2%) devices, including 458 malfunctions, 30 injuries, and 1 death. A total of 40 devices (5.8%) were recalled 113 times, primarily due to software issues.

CONCLUSIONS AND RELEVANCE: This cross-sectional study suggests that despite increasing clearance of AI/ML devices, standardized efficacy, safety, and risk assessment by the FDA are lacking. Dedicated regulatory pathways and postmarket surveillance of AI/ML safety events may address these challenges.

Для цитирования: Lin John C, Jain Bhav, Iyer Jay M, Rola Ishan, Srinivasan Anusha R, Kang Chaerim, Patel Heta, Parikh Ravi B. Benefit-Risk Reporting for FDA-Cleared Artificial Intelligence-Enabled Medical Devices.. JAMA health forum. 2025. DOI: 10.1001/jamahealthforum.2025.3351.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.