Randomised crossover trial comparing algorithms and averaging times for automatic oxygen control in preterm infants.

Schwarz Christoph E, Kreutzer Karen B, Langanky Lukas, Wolf Nicole S, Braun Wolfgang, O'Sullivan Marc Paul, Poets Christian F, Franz Axel R

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

Automatic control (SPOC) of the fraction of inspired oxygen (FiO2), based on continuous analysis of pulse oximeter saturation (SpO2), improves the proportion of time preterm infants spend within a specified SpO2-target range (Target%). We evaluated if a revised SPOC algorithm (SPOCnew, including an upper limit for FiO2) compared to both routine manual control (RMC) and the previously tested algorithm (SPOCold, unrestricted maximum FiO2) increases Target%, and evaluated the effect of the pulse oximeter's averaging time on controlling the SpO2 signal during SPOC periods.

DESIGN

Unblinded, randomised controlled crossover study comparing 2 SPOC algorithms and 2 SpO2 averaging times in random order: 12 hours SPOCnew and 12 hours SPOCold (averaging time 2 s or 8 s for 6 hours each) were compared with 6-hour RMC. A generated list of random numbers was used for allocation sequence.

SETTING

University-affiliated tertiary neonatal intensive care unit, Germany PATIENTS: Twenty-four infants on non-invasive respiratory support with FiO2 >0.21 were analysed (median gestational age at birth, birth weight and age at randomisation were 25.3 weeks, 585 g and 30 days).

MAIN OUTCOME MEASURE: Target%.

RESULTS

Mean (SD) [95% CI] Target% was 56% (9) [52, 59] for RMC versus 69% (9) [65, 72] for SPOCold_2s, 70% (7) [67, 73] for SPOCnew_2s, 71% (8) [68, 74] for SPOCold_8s and 72% (8) [69, 75] for SPOCnew_8s.

CONCLUSIONS

Irrespective of SpO2-averaging time, Target% was higher with both SPOC algorithms compared to RMC. Despite limiting the maximum FiO2, SPOCnew remained significantly better at maintaining SpO2 within target range compared to RMC.

TRIAL REGISTRATION: NCT03785899.

Для цитирования: Schwarz Christoph E, Kreutzer Karen B, Langanky Lukas, Wolf Nicole S, Braun Wolfgang, O'Sullivan Marc Paul, Poets Christian F, Franz Axel R. Randomised crossover trial comparing algorithms and averaging times for automatic oxygen control in preterm infants.. Archives of disease in childhood. Fetal and neonatal edition. 2021. DOI: 10.1136/archdischild-2021-322096.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.