MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jelectrocard.2020.02.005

Медицинская библиотека · Journal of electrocardiology · 2020

Comparison of electrocardiogram quality and clinical interpretations using prepositioned ECG electrodes and conventional individual electrodes.

Roy Sion K, Shah Sonia U, Villa-Lopez Eva, Murillo Mary, Arenas Nataly, Oshima Karin, Chang Ruey-Kang, Lauzon Marie, Guo Xiuqing, Pillutla Priya

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Efforts have been made to simplify and reduce technical errors, such as limb leads reversal and inaccurate chest leads placement, for the 12-lead ECG tests. We compared standard ECG using individual electrodes with a novel pre-positioned electrode system to determine equivalency.

METHODS

Subjects were recruited from the Emergency Department and cardiac lab of an acute care hospital in Los Angeles. First, subjects underwent a conventional 12-lead ECG using Philips PageWriter (clinical ECG). A second ECG was then performed using a novel system containing pre-positioned electrodes and a compact recorder (study ECG). All ECGs were reviewed by 3 blinded, board-certified adult cardiologists using 14 pre-specified ECG diagnostic categories to determine if the interpretations of clinical ECG and study ECG of the same patient were "equivalent". Majority rule was applied when there were discrepant interpretations among the 3 cardiologists.

RESULTS

One hundred subjects, ages 18 to 74 completed the study. With pre-positioned electrodes, the rate of "electrode fit" as judged by the research associates at the time of lead placement was 96.2%. We found that the study ECG system was equivalent (in clinical interpretation) to the clinical ECG system, with equivalency rate of 96% (95% confidence interval 92% to 100%) in "overall interpretation". The equivalence rate for the 14 ECG diagnostic categories ranged from 96% to 100%, with mean 99.2 ± 1.1%.

CONCLUSIONS

12-lead ECGs performed using single-piece, pre-positioned electrodes are clinically equivalent to those performed using 10 individually placed conventional electrodes. With 4 sizes for adults, the single-piece electrodes can fit 96% of the study patients.

Для цитирования: Roy Sion K, Shah Sonia U, Villa-Lopez Eva, Murillo Mary, Arenas Nataly, Oshima Karin, Chang Ruey-Kang, Lauzon Marie, Guo Xiuqing, Pillutla Priya. Comparison of electrocardiogram quality and clinical interpretations using prepositioned ECG electrodes and conventional individual electrodes.. Journal of electrocardiology. 2020. DOI: 10.1016/j.jelectrocard.2020.02.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.