Multicenter performance evaluation of the Elecsys HCV Duo immunoassay.

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Majchrzak Mario, Bronner Korbinian, Laperche Syria, Riester Elena, Bakker Ed, Bollhagen Ralf, Klinkicht Markus, Vermeulen Marion, Schmidt Michael

Аннотация

BACKGROUND: The diagnostic accuracy of the Elecsys® HCV Duo antigen-antibody combination immunoassay (Roche Diagnostics GmbH) was evaluated for the detection of hepatitis C virus (HCV) infection, versus commercially available comparators.

METHODS: This multicenter study (August 2020-March 2021) assessed the specificity of the Elecsys HCV Duo immunoassay and comparator assays in blood donor and routine clinical laboratory samples; sensitivity was determined in confirmed HCV-positive samples and seroconversion panels. The Elecsys HCV Duo immunoassay was compared with the Monolisa HCV Ag-Ab ULTRA V2, Murex HCV Ag/Ab Combination and ARCHITECT HCV Ag assays, as well as nucleic acid testing (NAT). The antibody (anti-HCV) module of the Elecsys HCV Duo immunoassay was compared with the Elecsys Anti-HCV II, Alinity s Anti-HCV, ARCHITECT Anti-HCV and RIBA HCV 3.0 SIA assays.

RESULTS: The specificity of the Elecsys HCV Duo immunoassay was 99.94% (95% confidence interval [CI], 99.89-99.97) and 99.92% (95% CI, 99.71-99.99) in blood donor (n = 20,634) and routine clinical laboratory samples (n = 2531), respectively. The specificity of the Elecsys HCV Duo immunoassay was similar or better than comparator assays. The sensitivity of the Elecsys HCV Duo immunoassay in confirmed HCV-positive samples (n = 257) was 99.6%. In seroconversion panels, the Elecsys HCV Duo immunoassay detected infections earlier (2.2-21.9 days) than all but one of the comparator assays and detected HCV 1.8 days later than NAT.

CONCLUSIONS: The Elecsys HCV Duo immunoassay shows high diagnostic accuracy, reduces the diagnostic window, and could be used when NAT is not possible.