Multicenter performance evaluation of the Elecsys HCV Duo immunoassay.
Аннотация
BACKGROUND
The diagnostic accuracy of the Elecsys® HCV Duo antigen-antibody combination immunoassay (Roche Diagnostics GmbH) was evaluated for the detection of hepatitis C virus (HCV) infection, versus commercially available comparators.
METHODS
This multicenter study (August 2020-March 2021) assessed the specificity of the Elecsys HCV Duo immunoassay and comparator assays in blood donor and routine clinical laboratory samples; sensitivity was determined in confirmed HCV-positive samples and seroconversion panels. The Elecsys HCV Duo immunoassay was compared with the Monolisa HCV Ag-Ab ULTRA V2, Murex HCV Ag/Ab Combination and ARCHITECT HCV Ag assays, as well as nucleic acid testing (NAT). The antibody (anti-HCV) module of the Elecsys HCV Duo immunoassay was compared with the Elecsys Anti-HCV II, Alinity s Anti-HCV, ARCHITECT Anti-HCV and RIBA HCV 3.0 SIA assays.
RESULTS
The specificity of the Elecsys HCV Duo immunoassay was 99.94% (95% confidence interval [CI], 99.89-99.97) and 99.92% (95% CI, 99.71-99.99) in blood donor (n = 20,634) and routine clinical laboratory samples (n = 2531), respectively. The specificity of the Elecsys HCV Duo immunoassay was similar or better than comparator assays. The sensitivity of the Elecsys HCV Duo immunoassay in confirmed HCV-positive samples (n = 257) was 99.6%. In seroconversion panels, the Elecsys HCV Duo immunoassay detected infections earlier (2.2-21.9 days) than all but one of the comparator assays and detected HCV 1.8 days later than NAT.
CONCLUSIONS
The Elecsys HCV Duo immunoassay shows high diagnostic accuracy, reduces the diagnostic window, and could be used when NAT is not possible.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.