m+Medelevator.RU
← Поиск публикаций
Clinical chemistry and laboratory medicine · 2020

Establishing the 99th percentile of a novel assay for high-sensitivity troponin I in a healthy blood donor population.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Odsæter Ingrid Hov, Grenne Bjørnar, Hov Gunhild Garmo, Laugsand Lars Erik, Wiseth Rune, Mikkelsen Gustav

Аннотация

BACKGROUND: The recommended cut-off of cardiac troponin (cTn) for the diagnosis of acute myocardial infarction (AMI) is the 99th percentile in a healthy reference population. We aimed to determine the 99th percentile of the novel ADVIA Centaur® High Sensitivity Troponin I assay (Siemens Healthcare Diagnostics) in fresh lithium heparin plasma samples from healthy blood donors.

METHODS: A total of 1000 apparently healthy blood donors were included. High-sensitivity (hs) cTnI, hs-cTnT, creatinine and N-terminal pro b-type natriuretic peptide (NT-proBNP) were measured in fresh lithium heparin plasma samples, and glycated hemoglobin (HbA1c) was measured in ethylenediaminetetraacetic acid (EDTA)-blood. The 99th percentile was estimated for the whole population, as well as for males and females separately.

RESULTS: For the total population the 99th percentile of ADVIA Centaur® High Sensitivity Troponin I was 96 (65-149) ng/L. The estimated value differed significantly from results published by others and was highly dependent on which values were considered statistical outliers.

CONCLUSIONS: The estimated 99th percentile for hs-cTnI in the population studied differed significantly from previously published results. There is a need for further specifications regarding how subjects used for estimating the 99th percentile of cTns in healthy populations should be recruited and how outlier values should be identified, as this can highly influence the diagnostic cut-off applied for AMI.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты