DOI 10.4103/jlp.jlp_71_17

Медицинская библиотека · Journal of laboratory physicians · 2018

Evaluation of new indigenous "point-of-care" ABO and Rh grouping device.

Tiwari Aseem Kumar, Setya Divya, Aggarwal Geet, Arora Dinesh, Dara Ravi C, Ratan Ankita, Bhardwaj Gunjan, Acharya Devi Prasad

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Erycard 2.0 is a "point-of-care" device that is primarily being used for patient blood grouping before transfusion.

MATERIALS AND

METHODS

Erycard 2.0 was compared with conventional slide technology for accuracy and time taken for ABO and Rh forward grouping result with column agglutination technology (CAT) being the gold standard. Erycard 2.0 as a device was also evaluated for its stability under different storage conditions and stability of result till 48 h. In addition, grouping of hemolyzed samples was also tested with Erycard 2.0. Ease of use of Erycard 2.0 was evaluated with a survey among paramedical staff.

RESULTS

Erycard 2.0 demonstrated 100% concordance with CAT as compared with slide technique (98.9%). Mean time taken per test by Erycard 2.0 and slide technique was 5.13 min and 1.7 min, respectively. After pretesting storage under different temperature and humidity conditions, Erycard 2.0 did not show any deviation from the result. The result did not change even after 48 h of testing and storage under room temperature. 100% concordance was recorded between pre- and post-hemolyzed blood grouping. Ease of use survey revealed that Erycard 2.0 was more acceptable to paramedical staff for its simplicity, objectivity, and performance than conventional slide technique.

CONCLUSION

Erycard 2.0 can be used as "point-of-care" device for blood donor screening for ABO and Rh blood group and can possibly replace conventional slide technique.

Для цитирования: Tiwari Aseem Kumar, Setya Divya, Aggarwal Geet, Arora Dinesh, Dara Ravi C, Ratan Ankita, Bhardwaj Gunjan, Acharya Devi Prasad. Evaluation of new indigenous "point-of-care" ABO and Rh grouping device.. Journal of laboratory physicians. 2018. DOI: 10.4103/jlp.jlp_71_17.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.