A uniform method for the simultaneous blood group phenotyping of Fya , Fyb , Jka , Jkb , S, s̅, P1, k applying lateral-flow technique.
Аннотация
BACKGROUND AND
OBJECTIVES
A lateral flow assay for simultaneous blood group typing of ABO, RhD, C, E, c, e, Cw and K with stable end-point and without centrifugation is in routine use since several years (MDmulticard® ). The typing of extended phenotype parameters belonging to the Duffy, Kidd, MNSs blood group systems and others, however, has not yet been demonstrated for this technique. Reliable detection of Fyx , a weak Fyb phenotype with a pronounced quantitative reduction of the number of Fyb antigens on the erythrocyte surface, remains a weakness of current serological blood grouping techniques.
MATERIAL AND
METHODS
The performance characteristics of the following reagents were evaluated in donor and patient samples in lateral flow technology (MDmulticard® ): Anti-Fya , -Fyb , -Jka , -Jkb , -S, -s̅, -P1 and -k. The sensitivity to detect Fyx was in addition evaluated with Fyx positive samples, which had been preselected by MALDI-TOF MS-based genotyping.
RESULTS
All results obtained with the MDmulticard® were in full accordance with those of the CE-certified reference products for all the eight reagent formulations used: Anti-Fya , -Fyb , -Jka , -Jkb , -S, -s̅, -P1 and -k. Also, all Fyx phenotypes of the selected population of 93 positive samples, originally identified by MALDI-TOF MS-based genotyping, were reliably detected by the lateral flow assay.
CONCLUSION
Extended phenotype blood group parameters, including the serologically challenging Fyx phenotype, can be determined simultaneously, rapidly and accurately using the lateral flow (MDmulticard® ) technology, even in cases when IgG class antibodies are the only source of diagnostic antibodies.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.