[Establishment of Internal Quality Control Method for Mixed and Split Samples by NAT in Blood Donor Screening].

Tang Xin-Yi, Ji Yun-Peng, Feng Chen-Chen

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To establish and evaluate the methods of internal quality control in blood donor screening by nucleic acid test (NAT).

METHODS

After HBV-DNA standard quality control (QC) sample (60 IU/ml) was diluted by pooling 6 samples, the concentration was 10.0 IU/ml, which was approach twice of the low limit. When the pooling result turned out reactive, the pooling samples need to be split into single sample to process. Meanwhile, the standard QC samples were tested as well. The same batch QC samples were tested 20 times respectively, calculate the mean (x̄), standard deviation (SD) and CV. Make Levey-Jennings QC curves by setting x̄±2SD as warning, x̄±3SD as rejected. The Levey-Jennings quality controls chart was mapped by using Microsoft Excel.

RESULTS

After 20 times test of mixed/split samples, the x̄±2SD were 33.03±1.47 and 30.08±0.98, the x̄±3SD were 33.03±2.20 and 30.08±1.47, the CV were 2.22% and 1.63%, respectively. The P value of t test was 0.08 and 0.17 respectively, there was no statistically significant difference between the 2 group.

CONCLUSION

When establish an internal QC system in the screening laboratory by nucleic acid testing, the concentration of the QC samples should be equal to normal specimens. This type of QC system may validate the extraction and amplification of the nucleic acid, and improve the stability of the test results.

Для цитирования: Tang Xin-Yi, Ji Yun-Peng, Feng Chen-Chen. [Establishment of Internal Quality Control Method for Mixed and Split Samples by NAT in Blood Donor Screening].. Zhongguo shi yan xue ye xue za zhi. 2016. DOI: 10.7534/j.issn.1009-2137.2016.03.052.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.