Quality evaluation of four hemoglobin screening methods in a blood donor setting along with their comparative cost analysis in an Indian scenario.

Tondon Rashmi, Verma Anupam, Pandey Prashant, Chaudhary Rajendra

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Despite the wide range of methods available for measurement of hemoglobin, no single technique has emerged as the most appropriate and ideal for a blood donation setup.

MATERIALS AND

METHODS

A prospective study utilizing 1014 blood samples was carried out in a blood donation setting for quality evaluation of four methods of hemoglobin estimation along with cost analysis: Hematology cell analyzer (reference), HCS, CuSO4 method and HemoCue.

RESULTS

Mean value of HemoCue (mean +/- SD = 14.7 +/- 1.49 g/dl) was higher by 0.24 compared to reference (mean +/- SD = 13.8 +/- 1.52 g/dl) but not statistically significant ( P > 0.05). HemoCue proved to be the best technique (sensitivity 99.4% and specificity 84.4%) whereas HCS was most subjective with 25.2% incorrect estimations. CuSO4 proved to be good with 7.9% false results. Comparative cost analysis of each method was calculated to be 35 INR/test for HemoCue, 0.76 INR /test for HCS and 0.06-0.08 INR /test for CuSO4.

CONCLUSION

CuSO4 method gives accurate results, if strict quality control is applied. HemoCue is too expensive to be used as a primary screening method in an economically restricted country like India.

Для цитирования: Tondon Rashmi, Verma Anupam, Pandey Prashant, Chaudhary Rajendra. Quality evaluation of four hemoglobin screening methods in a blood donor setting along with their comparative cost analysis in an Indian scenario.. Asian journal of transfusion science. 2009. DOI: 10.4103/0973-6247.53874.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.