DOI 10.5858/1999-123-0642-btahc
Blood transfusion-acquired hemoglobin C.
Аннотация
Unexpected and confusing laboratory test results can occur if a blood sample is inadvertently collected following a blood transfusion. A potential for transfusion-acquired hemoglobinopathy exists because heterozygous individuals show no significant abnormalities during the blood donor screening process. Such spurious results are infrequently reported in the medical literature. We report a case of hemoglobin C passively transferred during a red blood cell transfusion. The proper interpretation in our case was assisted by calculations comparing expected hemoglobin C concentration with the measured value. A review of the literature on transfusion-related preanalytic errors is provided.