DOI 10.1016/j.jacep.2017.06.022

Медицинская библиотека · JACC. Clinical electrophysiology · 2017

Safety and Efficacy of Multipoint Pacing in Cardiac Resynchronization Therapy: The MultiPoint Pacing Trial.

Niazi Imran, Baker James, Corbisiero Raffaele, Love Charles, Martin David, Sheppard Robert, Worley Seth J, Varma Niraj, Lee Kwangdeok, Tomassoni Gery, MPP Investigators

Сравнительное исследованиеСтатьяМногоцентровое исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVES

The MultiPoint Pacing (MPP) trial assessed the safety and efficacy of pacing 2 left ventricular sites with a quadripolar lead in patients with heart failure indicated for a CRT-D device.

BACKGROUND

Cardiac resynchronization therapy nonresponse is a complex problem where stimulation of multiple left ventricular sites may be a solution.

METHODS

Enrolled patients were indicated for a CRT-D system. Bi-ventricular (Bi-V) pacing was activated at implant. Three months later, clinical response was assessed and the patient was randomized (1:1) to receive Bi-V pacing or MPP. Patients were followed for 6 months post-randomization and clinical response was again assessed.

RESULTS

The CRT-D system was successfully implanted in 455 of 469 attempted implants (97%). A total of 381 patients were randomized to Bi-V or MPP at 3 months. The primary safety endpoint was met with freedom from system-related complications of 93.2%. The primary efficacy endpoint of the noninferiority comparison of nonresponder rates between the 2 arms was met. Patients randomized to MPP arm and programmed to pace from anatomically distant poles (MPP-AS) responded to therapy at significantly higher rates than MultiPoint pacing-other programmed settings (MPP-Other). Within this group, 87% were responders at 9 months, 100% designated as nonresponders at 3 months converted to responders at 9 months, and 54% experienced an incremental response compared to MPP-Other. Also within MPP-AS, 92% of patients with de novo CRT-D implant were classified as responders compared with patients with MPP-Other.

CONCLUSIONS

MPP is safe and effective for treating heart failure. The study met the pre-specified hypothesis that response to MPP is noninferior to Bi-V pacing with a quadripolar left ventricular lead. (MultiPoint Pacing IDE Study [MPP IDE]; NCT01786993).

Для цитирования: Niazi Imran, Baker James, Corbisiero Raffaele, Love Charles, Martin David, Sheppard Robert, Worley Seth J, Varma Niraj, Lee Kwangdeok, Tomassoni Gery, MPP Investigators. Safety and Efficacy of Multipoint Pacing in Cardiac Resynchronization Therapy: The MultiPoint Pacing Trial.. JACC. Clinical electrophysiology. 2017. DOI: 10.1016/j.jacep.2017.06.022.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.