Clinical evaluation of the BioFire FilmArray® BioThreat-E test for the diagnosis of Ebola Virus Disease in Guinea.

Gay-Andrieu Françoise, Magassouba N'Fally, Picot Valentina, Phillips Cynthia L, Peyrefitte Christophe N, Dacosta Brigitte, Doré Ahmadou, Kourouma Fode, Ligeon-Ligeonnet Véronique, Gauby Corentin, Longuet Christophe, Scullion Matt, Faye Ousmane, Machuron Jean Louis, Miller Mark

Оценочное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The recent West Africa Ebola outbreak highlighted the need to provide access to rapid, safe and reliable Ebola Virus Disease diagnostics.

OBJECTIVES

The objective of this field study was to assess the clinical performance of the FilmArray® BioThreat-E test for the detection of Ebola Zaïre virus in whole blood in symptomatic patients suspected of Ebola Virus Disease in Conakry (Guinea) from March to July 2015.

STUDY

DESIGN

The BioThreat-E test was compared to the two RT-PCRs, using serum, implemented at Donka Hospital in the emergency context: an in-house developed quantitative one-step RT-PCR adapted from the Weidmann technique, and the RealStar® Filovirus RT-PCR Kit 1.0 (Altona-Diagnostics). We also assessed the performance of this assay in noninvasive specimens (urine and saliva) to detect infected patients.

RESULTS

Of 135 patients enrolled and eligible for performance assessment on whole blood, the sensitivity was 95.7% [95% CI: 85.5-99.5] and specificity 100% [95% CI: 95.9-100]. Of the 37 symptomatic infected patients able to provide saliva and/or urine samples, 34 of the 35 saliva samples and all 3 of the urine samples were positive with the BioThreat-E test.

CONCLUSIONS

This study showed that the FilmArray BioThreat-E test performs comparably to conventional molecular tests under field conditions, providing results and interpretation in approximately 1h. Due to its operational characteristics, it can be easily deployed in the field during an epidemic and could also be a useful tool for post-outbreak surveillance.

Для цитирования: Gay-Andrieu Françoise, Magassouba N'Fally, Picot Valentina, Phillips Cynthia L, Peyrefitte Christophe N, Dacosta Brigitte, Doré Ahmadou, Kourouma Fode, Ligeon-Ligeonnet Véronique, Gauby Corentin, Longuet Christophe, Scullion Matt, Faye Ousmane, Machuron Jean Louis, Miller Mark. Clinical evaluation of the BioFire FilmArray® BioThreat-E test for the diagnosis of Ebola Virus Disease in Guinea.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2017. DOI: 10.1016/j.jcv.2017.04.015.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.