DOI 10.1016/j.radonc.2026.111801
Technical and biological validation of prostate apparent diffusion coefficient measurements on an MRI-linac.
Аннотация
PURPOSE: Quantitative MRI biomarkers require validation before being used to assess treatment response or inform biological image-guided adaptive radiotherapy on MRI-linacs. This work contributes to the technical validation of prostate apparent diffusion coefficient (ADC) measurements by evaluating phantom accuracy, intra-scanner repeatability, and inter-scanner reproducibility, and contributes to biological validation by evaluating correlations with histology.
METHODS
Phantom ADC was measured on a 1.5 T MRI-linac (MRL) and 1.5 T diagnostic MR (dMR). ADC in non-lesion tissue (ADCNL) and in the dominant intraprostatic lesion (ADCDIL) was measured twice in 16 patients on MRL and/or dMR before prostatectomy. Repeatability was quantified through Bland-Altman analysis and within-subject coefficients of variation (wCV). Reproducibility was quantified through Bland-Altman analysis and comparison with diffusion microstructural model simulations. Tissue samples were analysed using haematoxylin and eosin (H&E) and immunohistochemistry (IHC). Spearman's ρ quantified relationships between ADC and histology-derived epithelial (EF) and stroma (SF) fraction.
RESULTS
Phantom mean bias was 0.024 (MRL) and -0.019 (dMR) µm2/ms. Patient ADCs pooled across MRL and dMR exhibited excellent repeatability: wCVNL = 1.4%, wCVDIL = 4.6%. ADCNL was higher on MRL than dMR: mean bias [95% confidence interval] = 0.20 [0.14, 0.26] µm2/ms, consistent with simulations. Correlations between ADCDIL and histology-derived metrics were consistent in magnitude and direction for both scanners. Pooled data showed ADCDIL and EFH&E were negatively correlated (ρ = -0.69, p = 0.003); ADCDIL and SFH&E were positively correlated (ρ = 0.70, p = 0.003).
CONCLUSION
Systematic differences exist between MRI-linac and diagnostic MR measurements. ADCDIL from both systems showed similar correlations with tissue microstructure. This study contributes to the technical and biological validation of diagnostic MR and MRI-linac prostate ADC.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.