Real-world assessment of Multi-Frequency Bioelectrical Impedance Analysis (MFBIA) for measuring body composition in healthy physically active populations.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Potter Adam W, Ward Leigh C, Chapman Christopher L, Tharion William J, Looney David P, Friedl Karl E
Аннотация
BACKGROUND: Multi-frequency bioelectrical impedance analysis (MFBIA) methods offer reliable and moderately accurate estimates of body composition in tightly controlled conditions (prandial and hydration status, recent exercise, time of day).
OBJECTIVE: This study examined MFBIA reliability and validity in a real-world environment where these factors were not controlled.
METHODS: Regional and total body composition estimates by MFBIA (InBody 770) were compared to dual-energy X-ray absorptiometry (DXA) in 1000 healthy adults (667 men; 333 women), including fat mass (FM), percent body fat (%BF), fat-free mass (FFM), and visceral adipose tissue (VAT). In subsets, reliability was determined from duplicate MFBIA and DXA obtained within 1 week, and total body water (TBW) was compared to single-frequency BIA (SFBIA).
RESULTS: MFBIA demonstrated modest population-level agreement with DXA for total body FM (men, r = 0.93, bias -3.7 ± 2.6 kg; women, r = 0.96, bias, -1.9 ± 1.8 kg), %BF (men, r = 0.89, bias, -4.2 ± 3.0%; women, r = 0.92, bias, -2.8 ± 2.6%), and FFM (men, r = 0.95, bias, 3.4 ± 2.8 kg; women, r = 0.94, bias, 2.0 ± 2.2 kg). Regional correlations were highest for trunk FM (men, r = 0.92, CCC = 0.86; women r = 0.93, CCC = 0.93) and lowest for VAT (men, r = 0.74, CCC = 0.68; women, r = 0.74, CCC = 0.34). DXA and MFBIA regional and total assessments were highly reliable (DXA, ICC 0.990-0.998) and (MFBIA, ICC 0.987-0.995). TBW by MFBIA and SFBIA showed moderate agreement (men, r = 0.73, bias, -1.89 ± 3.31; women, r = 0.82, bias, -1.74 ± 2.01).
CONCLUSION: This MFBIA system was shown to have high retest reliability and, when compared to laboratory methods, provides a moderately accurate method for measuring TBW and body composition (except for VAT) in uncontrolled conditions.
OBJECTIVE: This study examined MFBIA reliability and validity in a real-world environment where these factors were not controlled.
METHODS: Regional and total body composition estimates by MFBIA (InBody 770) were compared to dual-energy X-ray absorptiometry (DXA) in 1000 healthy adults (667 men; 333 women), including fat mass (FM), percent body fat (%BF), fat-free mass (FFM), and visceral adipose tissue (VAT). In subsets, reliability was determined from duplicate MFBIA and DXA obtained within 1 week, and total body water (TBW) was compared to single-frequency BIA (SFBIA).
RESULTS: MFBIA demonstrated modest population-level agreement with DXA for total body FM (men, r = 0.93, bias -3.7 ± 2.6 kg; women, r = 0.96, bias, -1.9 ± 1.8 kg), %BF (men, r = 0.89, bias, -4.2 ± 3.0%; women, r = 0.92, bias, -2.8 ± 2.6%), and FFM (men, r = 0.95, bias, 3.4 ± 2.8 kg; women, r = 0.94, bias, 2.0 ± 2.2 kg). Regional correlations were highest for trunk FM (men, r = 0.92, CCC = 0.86; women r = 0.93, CCC = 0.93) and lowest for VAT (men, r = 0.74, CCC = 0.68; women, r = 0.74, CCC = 0.34). DXA and MFBIA regional and total assessments were highly reliable (DXA, ICC 0.990-0.998) and (MFBIA, ICC 0.987-0.995). TBW by MFBIA and SFBIA showed moderate agreement (men, r = 0.73, bias, -1.89 ± 3.31; women, r = 0.82, bias, -1.74 ± 2.01).
CONCLUSION: This MFBIA system was shown to have high retest reliability and, when compared to laboratory methods, provides a moderately accurate method for measuring TBW and body composition (except for VAT) in uncontrolled conditions.