m+Medelevator.RU
← Поиск публикаций
JACC. Cardiovascular imaging · 2025

General Purpose Deep Learning Attenuation Correction Improves Diagnostic Accuracy of SPECT MPI: A Multicenter Study.

СтатьяМногоцентровое исследованиеКлиническое исследование III фазыНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Shanbhag Aakash D, Miller Robert J H, Lemley Mark, Kavanagh Paul, Liang Joanna X, Marcinkiewicz Anna M, Builoff Valerie, Van Kriekinge Serge, Ruddy Terrence D, Fish Mathews B, Einstein Andrew J, Martins Monica, Halcox Julian P, Kaufmann Philipp A, Buckley Christopher, Bateman Timothy M, Berman Daniel S, Dey Damini, Slomka Piotr J

Аннотация

BACKGROUND: Single-photon emission computed tomography (SPECT) myocardial perfusion imaging (MPI) uses computed tomography (CT)-based attenuation correction (AC) to improve diagnostic accuracy. Deep learning (DL) has the potential to generate synthetic AC images, as an alternative to CT-based AC.

OBJECTIVES: This study evaluated whether DL-generated synthetic SPECT images could enhance accuracy of conventional SPECT MPI.

METHODS: Study investigators developed a DL model in a multicenter cohort of 4,894 patients from 4 sites to generate simulated SPECT AC images (DeepAC). The model was externally validated in 746 patients from 72 sites in a clinical trial (A Phase 3 Multicenter Study to Assess PET Imaging of Flurpiridaz F 18 Injection in Patients With CAD; NCT01347710) and in 320 patients from another external site. In the first external cohort, the study assessed the diagnostic accuracy for obstructive coronary artery disease (CAD)-defined as left main coronary artery stenosis ≥50% or ≥70% in other vessels-for total perfusion deficit (TPD). In the latter, the study completed change analysis and compared quantitative scores for AC, DeepAC, and nonattenuation correction (NC) with clinical scores.

RESULTS: In the first external cohort (mean age, 63 ± 9.5 years; 69.0% male), 206 patients (27.6%) had obstructive CAD. The area under the receiver-operating characteristic curve (AUC) of DeepAC TPD (0.77; 95% CI: 0.73-0.81) was higher than the NC TPD (AUC: 0.73; 95% CI: 0.69-0.77; P < 0.001). In the second external cohort, DeepAC quantitative scores had closer agreement with actual AC scores compared with NC.

CONCLUSIONS: In a multicenter external cohort, DeepAC improved prediction performance for obstructive CAD. This approach could enhance diagnostic accuracy in facilities using conventional SPECT systems without requiring additional equipment, imaging time, or radiation exposure.