Height-preserving sizing method for the new Artisse intrasaccular device based on early findings from 26 cases of intracranial aneurysm embolization.

Chew Han Seng, Al-Ali Samer, Butler Benjamin, Chavda Swarupsinh, Lamin Saleh

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: With the recent relaunch of Artisse 2.0, optimal sizing strategies for the device are still being refined. The official sizing chart covers only taller-than-wide aneurysms within a limited diameter range of 3.4-6 mm. To expand its applicability, we propose an intuitive sizing strategy emphasizing the importance of preserving the device's natural height.

MATERIALS AND

METHODS

Patients treated with the Artisse 2.0 between November 2022 and October 2024 were retrospectively analyzed.

RESULTS

Among 26 aneurysms treated over a 24-month period, 13 exhibited varying degrees of 'dog-ear' formation on post-deployment imaging. These cases showed greater width oversizing (+2.06 vs +1.29 mm, P<0.05) and height reduction (-1.83 vs -0.53 mm, P<0.05) compared with cases with complete neck coverage. Similarly, 9 out of the 16 aneurysms with at least one surveillance imaging failed to achieve complete angiographic occlusion. These cases were also associated with greater width oversizing (+1.72 vs +1.43 mm, P=0.47) and height reduction (-2.18 vs -0.81 mm, P=0.052).

CONCLUSION

In contrast to the Woven EndoBridge (WEB) device, excessive lateral compression and height reduction appear to undermine the stability of the Artisse 2.0. As a result, the device size for spherical or ovoid aneurysms can be determined using the '+1/-1' method, while the '+1/-0' approach is recommended for cylindrical aneurysms to improve neck coverage through height preservation. These proposed sizing strategies may serve as a practical interim solution until more clinical data and user experience become available.

Для цитирования: Chew Han Seng, Al-Ali Samer, Butler Benjamin, Chavda Swarupsinh, Lamin Saleh. Height-preserving sizing method for the new Artisse intrasaccular device based on early findings from 26 cases of intracranial aneurysm embolization.. Journal of neurointerventional surgery. 2026. DOI: 10.1136/jnis-2025-023082.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.