MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/rpd/ncag115

Медицинская библиотека · Radiation protection dosimetry · 2026

The impact of breast shielding on radiation dose during computed tomography procedures.

Albeshan Salman, Aloud Haila Nasser A

СтатьяПеревод в очереди

Аннотация

To quantify and compare breast radiation dose across CT protocols using shielding and automatic tube current modulation (ATCM) in single- and dual-energy scanning modes. Surface breast dose was measured using a calibrated pencil ionization chamber placed in a female anthropomorphic phantom. Scans were performed on a 128-slice CT system across multiple anatomical regions using manual mA, ATCM, lead shielding, and single- and dual-energy modes. In single-energy abdominal CT, shielding reduced measured surface breast dose by 12.2%, and ATCM provided an additional 38.2% reduction. Dual-energy CT resulted in higher measured surface breast doses than single-energy, but shielding reduced dose by approximately 62%. The greatest reductions were observed in head and neck (63.7%) and facial bone scans (89.5%). Lead shielding and ATCM effectively reduced surface breast radiation dose, particularly in regions dominated by scatter. Dual-energy imaging increased dose, while combining shielding with ATCM provided the greatest reduction.

Для цитирования: Albeshan Salman, Aloud Haila Nasser A. The impact of breast shielding on radiation dose during computed tomography procedures.. Radiation protection dosimetry. 2026. DOI: 10.1093/rpd/ncag115.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.