Bridging Gaps in Newborn Screening: A Comparative Study of Point-of-Care Testing Versus Traditional Reference Laboratory Method for Congenital Hypothyroidism.

Indriastuti Endah, Rangkuti Rahmah Yasinta, Fitriani Fatimah Nur, Siswanto Putri Alief, Njoto Edwin Nugroho, Haykal Muhammad Nazhif, Aulia Fauqa Arinil

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

OBJECTIVE

This study aimed to compare the reliability, agreement, and practical application (turnaround time) of a point-of-care testing (POCT) device using two different reagents versus the traditional reference laboratory method for thyroid-stimulating hormone (TSH) measurement in newborn congenital hypothyroidism screening.

METHODOLOGY: This prospective cross-sectional method-comparison study included 122 term newborns at a mother and child hospital in Surabaya, Indonesia. Heel-prick capillary blood samples were simultaneously analyzed using the Standard F-200 POCT device with two reagents (Group 1: n = 62, Standard™ F TSH-II; Group 2: n = 60, FastClear F TSHII) and compared to the standard dried blood spot method (GSP® Neonatal hTSH, PerkinElmer) as reference. Analyses included Pearson and Spearman correlations, linear regression, Bland-Altman plots, intraclass correlation coefficient (ICC), Cohen's Kappa, and mean difference tests.

RESULTS

Strong positive correlations were observed in both Group 1 (Pearson's r = 0.87, Spearman's ρ = 0.85; p <0.001) and Group 2 (Pearson's r = 0.88, Spearman's ρ = 0.91; p <0.001). Bland-Altman analysis showed minimal systematic bias with acceptable agreement between the two techniques. The ICC values indicated excellent reliability (Group 1: 0.89; Group 2: 0.91). High categorical agreement was confirmed by Cohen's Kappa values of 0.85 and 0.87 (p <0.001). No significant differences in mean TSH levels were found between the POCT and reference methods. The POCT turnaround time was 30 minutes, significantly shorter than the 15-22 days required for the traditional method.

CONCLUSION

POCT devices present a promising alternative to traditional reference screening method for congenital hypothyroidism.

Для цитирования: Indriastuti Endah, Rangkuti Rahmah Yasinta, Fitriani Fatimah Nur, Siswanto Putri Alief, Njoto Edwin Nugroho, Haykal Muhammad Nazhif, Aulia Fauqa Arinil. Bridging Gaps in Newborn Screening: A Comparative Study of Point-of-Care Testing Versus Traditional Reference Laboratory Method for Congenital Hypothyroidism.. Journal of the ASEAN Federation of Endocrine Societies. 2026. DOI: 10.15605/jafes.041.02.6233.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.