Evaluation of an automated workflow for TSN-based bounded-latency communication in ISO IEEE 11073 SDC-based medical networks.

Dohms Maja, Wickel Noah, Radermacher Klaus, Janß Armin

СтатьяПеревод в очереди

Аннотация

PURPOSE: Interoperable, vendor-independent communication between medical devices in operating rooms has been enabled by the ISO IEEE 11073 Service-oriented Device Connectivity (SDC) standard since 2019. However, mechanisms for predictable and deterministic transmission behavior have not yet been standardized. One approach to enable bounded-latency communication is the use of Time-Sensitive Networking (TSN), an extension of standard Ethernet that enables time-synchronized and controlled data transmission to achieve reliable and time-deterministic data delivery.

METHODS

We implemented and evaluated the integration of TSN into SDC-based medical networks to provide bounded-latency communication through time-division mechanisms. An automated process was developed to extract transmission requirements from SDC devices and generate TSN configuration files for network deployment. The evaluation was conducted in a prototype SDC network under realistic traffic conditions. Furthermore, an SDC library designed with the aim of enabling faster data transmission via QUIC was used in the evaluation.

RESULTS

Automated TSN configuration is technically feasible and can improve bounded-latency communication in busy networks. With appropriately dimensioned schedules, the maximum SDC latency remains within the same range under high background traffic as under unloaded conditions. However, improperly dimensioned transmission windows can cause instability, connection loss and latency peaks of up to four times the planned transmission time.

CONCLUSION

SDC with TSN offers a promising approach for achieving bounded-latency communication in interoperable medical networks. Further optimization is required to eliminate latency peaks and meet the timing requirements of safety-critical medical applications.

Для цитирования: Dohms Maja, Wickel Noah, Radermacher Klaus, Janß Armin. Evaluation of an automated workflow for TSN-based bounded-latency communication in ISO IEEE 11073 SDC-based medical networks.. International journal of computer assisted radiology and surgery. 2026. DOI: 10.1007/s11548-026-03801-1.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.