DOI 10.1111/pace.70454

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2026

Lead-to-Vein Ratio and Long-Term Vascular Safety After Transvenous Pacing in Children Weighing ≤15 Kg.

Ortega Michel Cabrera, Novoa Juan Carlos Ramiro, Chirino Osmin Castañeda, Morejón Adel Eladio González, Ricardo Giselle Serrano, Hernández Yoel Coto, Gonzalez Mabel Elena Dominguez, Pérez Liz Coto, Fernández-Cisnal Agustín

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Transvenous pacing in low-weight children is technically feasible, but its clinical use depends on long-term venous patency, system performance, and ventricular function. We aimed to describe imaging-defined venous abnormalities and clinically overt vascular events after first transvenous pacemaker implantation and to explore whether the lead-to-vein ratio provides anatomical information beyond body weight.

METHODS

This prospective observational cohort included 35 children weighing ≤15 kg undergoing first transvenous pacemaker implantation between 2010 and 2020. The primary vascular endpoint was imaging-defined venous thrombosis, occlusion, or stenosis ≥50%; clinical impact was recorded separately. Device-related safety, minimum venous diameter, lead-to-vein ratio, ventricular function, synchrony, and paced QRS duration were analyzed using non-parametric, exact, and exploratory univariable tests.

RESULTS

Median follow-up was 6.7 [4.5-8.2] years. Eleven patients weighed <10 kg and 3 weighed <5 kg. The imaging-defined vascular endpoint occurred in 6/35 patients; 5 were asymptomatic surveillance findings and 1 was clinically symptomatic. The lead-to-vein ratio was higher in patients with the endpoint than in those without events (0.59 [0.55-0.63] vs. 0.47 [0.43-0.51]; p = 0.003). A ratio >0.55 was associated with more frequent imaging-defined endpoints (4/8 vs. 2/27; odds ratio 12.5; 95% confidence interval 1.7-92.3). Left ventricular ejection fraction remained preserved.

CONCLUSION

During the first transvenous system lifespan, imaging-defined venous abnormalities were relatively frequent but mostly asymptomatic, whereas clinically overt vascular complications were uncommon. The lead-to-vein ratio may support anatomical pre-implant risk stratification, although multicenter validation and follow-up beyond generator replacement are required.

Для цитирования: Ortega Michel Cabrera, Novoa Juan Carlos Ramiro, Chirino Osmin Castañeda, Morejón Adel Eladio González, Ricardo Giselle Serrano, Hernández Yoel Coto, Gonzalez Mabel Elena Dominguez, Pérez Liz Coto, Fernández-Cisnal Agustín. Lead-to-Vein Ratio and Long-Term Vascular Safety After Transvenous Pacing in Children Weighing ≤15 Kg.. Pacing and clinical electrophysiology : PACE. 2026. DOI: 10.1111/pace.70454.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.