DOI 10.1007/s00066-026-02599-w

Медицинская библиотека · Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] · 2026

Development of a prospective clinical workflow for proton beam therapy of ventricular tachycardia.

Cherit Hernández Ariadna, Fuss Martina C, Grehn Melanie, Trojani Valeria, Schmitzer Claus, Gruber-Hammer Günter, Xie Jingyang, Georg Dietmar, Fiedler Lukas, Fossati Piero, Blanck Oliver, Stock Markus

СтатьяПеревод в очереди

Аннотация

PURPOSE: This study presents a prospective preclinical workflow for proton beam therapy of refractory ventricular tachycardia (VT), covering target transfer, treatment planning (TP), and beam gating.

METHODS

Photon-based stereotactic arrhythmia radioablation (STAR) protocols were adapted to proton beam therapy, encompassing (i) target transfer from electroanatomical mapping (EAM) onto planning computed tomography (CT), (ii) simulation of three TP cases from the Standardized Treatment and Outcome Platform for Stereotactic Therapy of Re-Entrant Tachycardia by a Multidisciplinary (STOPSTORM.eu) benchmark, and (iii) evaluation of trigger events and delays for synchrotron beam delivery control using gating.

RESULTS

CARDIO-RT enabled electroanatomical mapping to computed tomography (EAM-to-CT) target transfer; two experienced cardiologists qualitatively judged the transferred targets as clinically acceptable by visual inspection. Treatment plans demonstrated effective organ at risk (OAR) sparing and clinical adequacy. The planning target volume (PTV) D95% ranged from 23.3 to 23.9 Gy relative biological effectiveness (RBE) and the clinical target volume electrophysiology (CTV [EP]) D95% from 30.9 to 31.6 Gy (RBE); underdosage was permitted where anatomical compromise was unavoidable. No CTV(EP) hotspot exceeded 33.5 Gy (RBE). Optical trigger signals are supported for beam gating; the total cardiac gating chain latency was below 30 ms.

CONCLUSION

This technical feasibility study demonstrates a prospective preclinical workflow for proton beam STAR, covering EAM-to-CT target transfer, TP across three benchmark clinical cases, and accelerator-side optical gating integration at a synchrotron-based facility. The workflow demonstrates technical feasibility; however, full clinical translation still requires patient positioning and verification, motion characterization, and end-to-end validation. Future work will employ a dynamic anthropomorphic phantom to characterize motion-induced dosimetric changes and assess plan robustness for proton and carbon ion beam delivery.

Для цитирования: Cherit Hernández Ariadna, Fuss Martina C, Grehn Melanie, Trojani Valeria, Schmitzer Claus, Gruber-Hammer Günter, Xie Jingyang, Georg Dietmar, Fiedler Lukas, Fossati Piero, Blanck Oliver, Stock Markus. Development of a prospective clinical workflow for proton beam therapy of ventricular tachycardia.. Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al]. 2026. DOI: 10.1007/s00066-026-02599-w.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.