Automatic Respiratory and Bulk Patient Motion Corrected 3D Fetal MRI.
Аннотация
PURPOSE: To develop a framework (ACROBATIC) for correcting motion in 3D radial fetal MRI.
METHODS
Data were simulated (N = 200) and acquired in utero (N = 11, gestational age: 32 ± 2 weeks). Motion due to maternal respiration was estimated by extracting a self-gating signal and applying focused navigation. Bulk motion was estimated by splitting the acquisition into sequential bins, reconstructing 3D volumes and applying rigid image registration. These combined motion estimates were used to correct k-space. Self-gating signals were compared to ground truth in simulations and an external sensor in utero. The cumulative position error (CPE) measured the accuracy of motion estimations relative to ground truth in simulations and relative sharpness measured the corresponding impact on image quality for both simulations and in utero data. An expert reviewer performed a blinded ranking of in utero images including uncorrected and corrected data.
RESULTS
Self-gating signals correlated strongly with ground truth for simulations (R = 0.97 ± 0.01) and a external sensor for in utero data (R = 0.75 ± 0.23). CPE decreased significantly using ACROBATIC (uncorrected: 13.33[12.73-14.05], corrected: 2.20[1.80-2.61]). Relative image sharpness increased with ACROBATIC for both simulated (4.51[2.89-5.79]) and in utero data (1.12[0.77-1.44]) consistent with expert ranking where ACROBATIC images were given the best rank in the majority of cases.
CONCLUSION
ACROBATIC enables motion correction in 3D radial fetal MRI. Correction of displacement due to maternal respiration and bulk motion results in improved image quality in simulations and in utero. Comparison to 2D frameworks are now warranted to establish the added diagnostic value of this approach.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.