Automated adaptive detection and reconstruction of quiescent cardiac phases in free-running whole-heart acquisitions using Synchronicity Maps from PHysiological mOtioN In Cine (SYMPHONIC) MRI.

Bongiolatti Giulia M C Rossi, Masala Nemanja, Bastiaansen Jessica A M, Yerly Jérôme, Prša Milan, Rutz Tobias, Tenisch Estelle, Si-Mohamed Salim, Stuber Matthias, Roy Christopher W

СтатьяПеревод в очереди

Аннотация

PURPOSE: To reconstruct whole-heart images from free-running acquisitions through automated selection of data acceptance windows (ES: end-systole, MD: mid-diastole, ED: end-diastole) that account for heart rate variability (HRV).

METHODS

SYMPHONIC was developed and validated in simulated (N = 1000) and volunteer (N = 14) data. To validate SYMPHONIC, the position of the detected acceptance windows, total duration, and resulting ventricular volume were compared to the simulated ground truth to establish metrics for temporal error, quiescent interval duration, and volumetric error, respectively. SYMPHONIC MD images and those using manually defined acceptance windows with fixed (MANUALFIXED) or adaptive (MANUALADAPT) width were compared by measuring vessel sharpness (VS). The impact of HRV was assessed in patients (N = 6).

RESULTS

Mean temporal error was larger for MD than for ED and ED in both simulations and volunteers. Mean volumetric errors were comparable. Interval duration differed for ES (p = 0.04) and ED (p < 10-3), but not for MD (p = 0.08). In simulations, SYMPHONIC and MANUALADAPT provided consistent VS for increasing HRV, while VS decreased for MANUALFIXED. In volunteers, VS differed between MANUALADAPT and MANUALFIXED (p < 0.01), but not between SYMPHONIC and MANUALADAPT (p = 0.03) or MANUALFIXED (p = 0.42).

CONCLUSION

SYMPHONIC accurately detected quiescent cardiac phases in free-running data and resulted in high-quality whole-heart images despite the presence of HRV.

Для цитирования: Bongiolatti Giulia M C Rossi, Masala Nemanja, Bastiaansen Jessica A M, Yerly Jérôme, Prša Milan, Rutz Tobias, Tenisch Estelle, Si-Mohamed Salim, Stuber Matthias, Roy Christopher W. Automated adaptive detection and reconstruction of quiescent cardiac phases in free-running whole-heart acquisitions using Synchronicity Maps from PHysiological mOtioN In Cine (SYMPHONIC) MRI.. Magma (New York, N.Y.). 2025. DOI: 10.1007/s10334-025-01289-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.