Comparison between fast-interrupted steady-state (FISS) and rapid water-excitation pulses for fat signal suppression in free-running whole-heart MRI at 1.5 T.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Safarkhanlo Yasaman, Yerly Jérôme, Falcão Mariana B L, Mackowiak Adèle L C, Piccini Davide, Stuber Matthias, Jung Bernd, Gräni Christoph, Bastiaansen Jessica A M
Аннотация
BACKGROUND: Free-running whole-heart MRI using balanced steady-state free precession (bSSFP) sequences offer high SNR and myocardial tissue contrast. However, an inadequate fat signal suppression may introduce artifacts and is particularly challenging with non-Cartesian readouts. The aim of this study was to evaluate different fat-signal suppression methods for whole-heart free-running MRI at 1.5 T using numerical simulations, phantom, and cardiac MRI experiments without the use of contrast agents.
METHODS: Binomial off-resonant rectangular (BORR), lipid insensitive binomial off-resonant RF excitation (LIBRE), and lipid insensitive binomial off-resonant (LIBOR) pulses were implemented within a 3D radial bSSFP sequence. Their pulse parameters were optimized for fat signal suppression at 1.5 T using simulations and phantom experiments. Optimized protocols, along with a free-running fast interrupted steady-state (FISS) and non-fat suppressed bSSFP sequence, were used to acquire phantom and cardiac data in five volunteers. SAR values were recorded. The SNR and CNRWater-Fat were measured in phantom data, while SNR and CNRBlood-Myocardium were quantified in volunteers using reconstruction without motion correction. Motion-resolved reconstructions were used for qualitative assessments. Statistical differences were analyzed using one-way ANOVA.
RESULTS: LIBOR had the highest CNRWater-Fat (276.8 ± 2.5) in phantoms, followed by LIBRE (268.1 ± 2.6), BORR (249.9 ± 2.2), and FISS (212.7 ± 2.7), though these differences were not statistically significant (p > 0.05). In volunteers, BORR had the highest SNR in the ventricular blood pool (17.0 ± 1.5), and LIBRE had the highest CNRBlood-Fat (29.4 ± 9.3). FISS had the highest CNRBlood-Myocardium (29.0 ± 8.9), but the differences were not significant (p > 0.05). Motion-resolved cardiac imaging showed comparable quality across all fat-suppressed sequences, with no significant streaking artifacts observed. Free-running bSSFP with LIBOR required the lowest SAR, up to a sixfold decrease compared with FISS.
CONCLUSION: The tested sequences performed similarly in SNR and CNR but LIBOR offered the lowest SAR, making it a promising candidate for applications where RF energy deposition is a concern.
METHODS: Binomial off-resonant rectangular (BORR), lipid insensitive binomial off-resonant RF excitation (LIBRE), and lipid insensitive binomial off-resonant (LIBOR) pulses were implemented within a 3D radial bSSFP sequence. Their pulse parameters were optimized for fat signal suppression at 1.5 T using simulations and phantom experiments. Optimized protocols, along with a free-running fast interrupted steady-state (FISS) and non-fat suppressed bSSFP sequence, were used to acquire phantom and cardiac data in five volunteers. SAR values were recorded. The SNR and CNRWater-Fat were measured in phantom data, while SNR and CNRBlood-Myocardium were quantified in volunteers using reconstruction without motion correction. Motion-resolved reconstructions were used for qualitative assessments. Statistical differences were analyzed using one-way ANOVA.
RESULTS: LIBOR had the highest CNRWater-Fat (276.8 ± 2.5) in phantoms, followed by LIBRE (268.1 ± 2.6), BORR (249.9 ± 2.2), and FISS (212.7 ± 2.7), though these differences were not statistically significant (p > 0.05). In volunteers, BORR had the highest SNR in the ventricular blood pool (17.0 ± 1.5), and LIBRE had the highest CNRBlood-Fat (29.4 ± 9.3). FISS had the highest CNRBlood-Myocardium (29.0 ± 8.9), but the differences were not significant (p > 0.05). Motion-resolved cardiac imaging showed comparable quality across all fat-suppressed sequences, with no significant streaking artifacts observed. Free-running bSSFP with LIBOR required the lowest SAR, up to a sixfold decrease compared with FISS.
CONCLUSION: The tested sequences performed similarly in SNR and CNR but LIBOR offered the lowest SAR, making it a promising candidate for applications where RF energy deposition is a concern.