Transit-guided radiation therapy: a novel patient monitoring approach.

Latorre-Musoll Artur, Oses Gabriela, Antelo Gabriela, Serrano-Rueda Sergi, Mollà Meritxell, Sempau Josep, Jornet Núria

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: Transit-Guided Radiation Therapy (TGRT) is a novel technique that uses the transit portal images (TPIs) acquired with Electronic Portal Image Devices (EPID) to quantify patient position errors during the treatment. It has been validated using anthropomorphic phantoms but a validation in a clinical setting was lacking. A pilot clinical study is presented to confirm our previous results.

MATERIALS AND

METHODS

A prospective study was conducted between June and December 2022 with patients who received whole-brain or breast radiotherapy treatments. The selected treatments were composed of radiation fields using skin-flash, where the body contour projected a sharp edge on the EPID which has been used as a surrogate of the true patient position. Daily imaging procedures were applied as scheduled before running the one- and two-parameter model (1PM and 2PM) of the TGRT formalism on the acquired TPIs to independently estimate the patient position errors.

RESULTS

43 patients and 1015 TPIs have been assessed. The 2PM showed a better correlation with the true position errors (R2 = 0.76 vs. 0.73), a lower detection threshold (0.77 mm vs. 1.24 mm), and a lower overcorrection risk above the detection threshold (7.0 % vs. 11.1 %) than the 1PM. Overall, the 2PM would have significantly reduced the true position errors by a factor of 0.58 (0.49 - 1.27) (p < 0.0001).

CONCLUSION

The TGRT technique has confirmed the ability to reduce the position errors in a clinical setting, demonstrating the potential to enhance the patient position monitoring without increasing treatment time or patient dose.

Для цитирования: Latorre-Musoll Artur, Oses Gabriela, Antelo Gabriela, Serrano-Rueda Sergi, Mollà Meritxell, Sempau Josep, Jornet Núria. Transit-guided radiation therapy: a novel patient monitoring approach.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2024. DOI: 10.1016/j.radonc.2024.110580.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.