Diagnostic performances of clinical laboratory tests using Triton X-100 to reduce the biohazard associated with routine testing of Ebola virus-infected patients.

Tempestilli Massimo, Pucci Luigia, Notari Stefania, Di Caro Antonino, Castilletti Concetta, Rivelli Maria Rosaria, Agrati Chiara, Pucillo Leopoldo Paolo

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Ebola virus, an enveloped virus, is the cause of the largest and most complex Ebola virus disease (EVD) outbreak in West Africa. Blood or body fluids of an infected person may represent a biohazard to laboratory workers. Laboratory tests of virus containing specimens should be conducted in referral centres at biosafety level 4, but based on the severity of clinical symptoms, basic laboratories might be required to execute urgent tests for patients suspected of EVD. The aim of this work was to compare the analytical performances of laboratory tests when Triton X-100, a chemical agent able to inactivate other enveloped viruses, was added to specimens.

METHODS

Results of clinical chemistry, coagulation and haematology parameters on samples before and after the addition of 0.1% (final concentration) of Triton X-100 and 1 h of incubation at room temperature were compared.

RESULTS

Overall, results showed very good agreement by all statistical analyses. Triton X-100 at 0.1% did not significantly affect the results for the majority of the analytes tested.

CONCLUSIONS

Triton X-100 at 0.1% can be used to reduce the biohazard in performing laboratory tests on samples from patients with EVD without affecting clinical decisions.

Для цитирования: Tempestilli Massimo, Pucci Luigia, Notari Stefania, Di Caro Antonino, Castilletti Concetta, Rivelli Maria Rosaria, Agrati Chiara, Pucillo Leopoldo Paolo. Diagnostic performances of clinical laboratory tests using Triton X-100 to reduce the biohazard associated with routine testing of Ebola virus-infected patients.. Clinical chemistry and laboratory medicine. 2015. DOI: 10.1515/cclm-2015-0119.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.