[Efficacy of L-shaped Shielding in Interventional Radiology by Transradial Approach and Consideration of Methods for Appropriate Use].

Аннотация на английскомСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Yamashita Katsunori, Ueno Hiroyuki, Haba Tomonobu, Ito Yusuke, Matsunaga Yuta, Hirosawa Ayaka, Ishizaki Souichirou, Mizui Masato, Osawa Mitsuharu, Nomura Takayuki, Mogi Satoshi

Аннотация

The present study investigated how effective an L-shaped shield was, depending on its position, in reducing a doctor's exposure to radiation during catheterization to access the transradial approach (TRA). The shield's effectiveness was evaluated by measuring the air kerma where the doctor stood under four conditions: with and without the shield, and with and without the shield in conjunction with conventional protection. To enable the shield to be positioned correctly in clinical practice, an illustrated instruction decal affixable to the shield's doctor-facing surface was produced, and the effectiveness of the decal was verified by means of a crossover test in which, as subjects of the study, different nurses set up the shield with and without the decal affixed to it. In the test, in which a human body phantom was used, the C-arm set at the PA angle, and the shield positioned 10 cm from the axilla of the phantom, the shield's effectiveness at 100 cm, 130 cm, and 160 cm above the floor where the doctor stood was 55%, 77%, and 47%, respectively. The effectiveness increased when the shield was positioned closer to the axilla. A significant difference in the positioning of the shield by the subjects was observed depending on whether or not the decal was affixed ( p<0.05, Wilcoxon signed-rank test), indicating that the use of the decal improved the positioning. It was concluded that, positioned correctly, the shield could effectively reduce the doctor's exposure to radiation during TRA.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты