Cyclosporin A Monitoring in Allogeneic Hematopoietic Stem-Cell Transplant Recipients: A Comparison Between Siemens Viva-ProE System and LC-MS/MS.

Lin Rongqi, Fan Xianghui, Li Xueyong, Lai Guixiang, Chen Ming, Huang Yingbin, Fu Qinghui, Lai Kaixian, Cheng Yu, Wu Xuemei, Qiu Hongqiang, Chen Yisheng

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Allogeneic hematopoietic stem-cell transplantation is a critical treatment for blood-related diseases often complicated by graft-versus-host disease and managed using cyclosporin A (CsA). Therapeutic drug monitoring is essential due to the narrow therapeutic window of CsA.

METHODS

In this study, a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the quantification of whole-blood CsA concentrations was developed and validated. CsA concentration in allogeneic hematopoietic stem-cell transplantation patients was measured using the newly established LC-MS/MS and the enzyme multiplied immunoassay technique-based Siemens viva-ProE system. The results from both methods were statistically analyzed across different concentration ranges to evaluate method agreement, correlation, and clinical implications.

RESULTS

The established LC-MS/MS method demonstrated excellent linearity (r 2 = 0.9983) within the CsA concentration range of 10-1000 ng/mL, with the validation parameters meeting the bioanalytical method validation criteria. Analysis of 161 samples revealed a strong correlation (r = 0.9908) between the methods; however, Siemens viva-ProE system overestimated CsA levels by an average of 8.7%, with a greater bias at higher concentrations. Correction through Deming regression significantly improved the agreement at therapeutically relevant thresholds.

CONCLUSIONS

The developed LC-MS/MS assay demonstrated acceptable performance in measuring CsA concentration in whole blood. The comparison between the 2 methods highlights the need for method specification in therapeutic drug monitoring and provides regression equations to enhance the monitoring accuracy in clinical practice.

Для цитирования: Lin Rongqi, Fan Xianghui, Li Xueyong, Lai Guixiang, Chen Ming, Huang Yingbin, Fu Qinghui, Lai Kaixian, Cheng Yu, Wu Xuemei, Qiu Hongqiang, Chen Yisheng. Cyclosporin A Monitoring in Allogeneic Hematopoietic Stem-Cell Transplant Recipients: A Comparison Between Siemens Viva-ProE System and LC-MS/MS.. Therapeutic drug monitoring. 2025. DOI: 10.1097/ftd.0000000000001390.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.